Performance Quick Check
Keysight
N4374B Lightwave Component Analyzer, User’s Guide
39
• Though please
note that this
trace is also
within
specifications.
• If there are
problems with the
electrical
calibration of the
network analyzer
- particularly if
the wrong
calibration planes
have been used -
the trace may look
like this, with
high frequency
ripples.
If the results differ
from the expected values
• Inspect the electrical connectors on the N4374B
Lightwave Component Analyzer and the “Flexible Cables
for Test Ports” visually.
If electrical connectors seems to be damaged contact
Keysight for repair or replacement.
• Clean the electrical connectors on the N4374B Lightwave
Component Analyzer and the “Flexible Cables for Test
Ports.
Perform a new electrical calibration and retest.
If the problem persists, contact Keysight for support.
If you have option 102 (Dual wavelength 1310nm, 1550nm), and
you are testing at both wavelengths, repeat the performance
quick check for the second (1310nm or 1550nm).