Sensor calibration tables
Sensor calibration tables store the measurement calibration factors for
thermistor sensors in the N432A. They are used to correct measurement
results.
Use sensor calibration tables when you want to make measurements over
a range of signal frequencies.
The N432A is capable of storing 20 sensor calibration tables, each
containing up to 80 frequency points. The N432A is supplied with a set of
two predefined sensor calibration tables plus a “100%” default table. The
data in these tables is based on statistical averages for a range of Keysight
thermistor sensors. Your own sensor will most likely differ from the
typical to some degree. If you require best accuracy, create a custom table
for each sensor you use as described in
calibration tables” on page 73
Selecting a sensor calibration table
You can select a calibration factor table from
>
>
.
The State column indicates if any calibration factor tables are currently
selected. The
Sensor Tbls
screen is shown in
.
N O T E
When no sensor tables are selected and the
display mode is chosen, the
calibration factor used for the measurement is displayed in the upper window.
Single Numeric
Tables
Sensor Cal Tables
N O T E
You need to enable the calibration factor table prior to configuring the measurement
frequency value in the channel setup screen.
N O T E
You can also access the sensor calibration tables by pressing
>
and
using the up/down arrow key to highlight the
CF Table
setting field and pressing
to
display the tables.
Offsets
Содержание N432A
Страница 2: ......
Страница 3: ......
Страница 20: ...XX N432A User s Guide ...
Страница 22: ...XXII N432A User s Guide ...
Страница 55: ...Operating Information 2 N432A User s Guide 33 4 Press to exit the channel setup screen ...
Страница 84: ...2 Operating Information 62 N432A User s Guide Figure 2 29 Min Hold indicated in full in the expanded window ...
Страница 90: ...2 Operating Information 68 N432A User s Guide Figure 2 36 Recall confirmation pop up dialog ...
Страница 118: ...2 Operating Information 96 N432A User s Guide ...
Страница 140: ...3 Maintenance 118 N432A User s Guide ...
Страница 152: ...4 Specifications and Characteristics 130 N432A User s Guide ...
Страница 153: ......