2
Operating Information
60
N432A User’s Guide
Max Hold/Min Hold
The max hold/min hold setting on the measurement setup window can be
configured on the front panel or via SCPI.
Min hold is the minimum of all measurements since the start time. The
min hold reading will be updated as a new minimum reading appears.
Max hold is the maximum of all measurements since the start time. The
max hold reading will be updated as a new maximum reading appears.
The max hold/min hold reading are resettable by pressing
.
Figure 2-26
Max hold/min hold measurement performed on the ‘HOLD’ block
Procedure
Use the following steps to set max hold/min hold.
1
Press
to access the measurement setup screen.
2
Use the up/down arrow key to highlight the
Hold
setting field.
3
Press
to display the
Hold
pop- up list, and use the up/down
arrow key to highlight
Min
or
Max
.
SENSe1:
Input from
the SENSe1
block
:FEED
FEED1
FEED2
:MATH
“A”
|
”B”
“A-A”
|
“A/A“
“B-A”
|
“B/A“
“A-B”
|
“A/B“
“B-B”
|
“B/B“
A
B
:GAIN
:HOLD
:REL
CALCulate Block
Max hold/
Min hold
P
avg
Содержание N432A
Страница 2: ......
Страница 3: ......
Страница 20: ...XX N432A User s Guide ...
Страница 22: ...XXII N432A User s Guide ...
Страница 55: ...Operating Information 2 N432A User s Guide 33 4 Press to exit the channel setup screen ...
Страница 84: ...2 Operating Information 62 N432A User s Guide Figure 2 29 Min Hold indicated in full in the expanded window ...
Страница 90: ...2 Operating Information 68 N432A User s Guide Figure 2 36 Recall confirmation pop up dialog ...
Страница 118: ...2 Operating Information 96 N432A User s Guide ...
Страница 140: ...3 Maintenance 118 N432A User s Guide ...
Страница 152: ...4 Specifications and Characteristics 130 N432A User s Guide ...
Страница 153: ......