Operating Information
42
N432A User’s Guide
Setting the channel offset
This gain or loss is applied to the measured power before any
mathematical functions, display offsets, or relative functions are included.
The offset is entered in dB within the range of –100 dB to +100 dB.
The equivalent range in % is from 0.0000000001% to 10000000000%.
A positive value compensates for a loss, while a negative value
compensates for a gain.
Procedure
Use the following steps to set a channel offset.
1
Press
to access the channel setup screen.
2
Press
to display the channel offset setup screen.
3
Use the up/down arrow key to highlight the
Offset
setting field.
4
Press
to select the
Offset
setting field.
Figure 2-9
Setting the channel offset
5
Press the right arrow key to highlight the
Offset
value field and press
to display the Offset pop- up. Use the numeric keypad to enter
the required value.
6
Confirm your entry by pressing
.
7
Press
to exit the offset setup screen. If either a channel or a
display offset is set, the
Ofs
indicator is displayed.
Offsets
Offset value field
Offset setting field
dB
Содержание N432A
Страница 2: ......
Страница 3: ......
Страница 20: ...XX N432A User s Guide ...
Страница 22: ...XXII N432A User s Guide ...
Страница 55: ...Operating Information 2 N432A User s Guide 33 4 Press to exit the channel setup screen ...
Страница 84: ...2 Operating Information 62 N432A User s Guide Figure 2 29 Min Hold indicated in full in the expanded window ...
Страница 90: ...2 Operating Information 68 N432A User s Guide Figure 2 36 Recall confirmation pop up dialog ...
Страница 118: ...2 Operating Information 96 N432A User s Guide ...
Страница 140: ...3 Maintenance 118 N432A User s Guide ...
Страница 152: ...4 Specifications and Characteristics 130 N432A User s Guide ...
Страница 153: ......