2
Operating Information
86
N432A User’s Guide
Figure 2-46
Connection diagram for VSWR measurement
23
Preset the power meter- under- test. Ensure that its oscillator/power
reference is turned off.
24
Set the DVMs to measure microvolts.
25
Round off the value of DVM 1 to two decimal places. Record this value
as V
0
. Observe the reading on DVM 2 and record the value as V
COMP0
.
26
Turn on the oscillator/power reference of the power meter- under- test.
27
Round off the value of DVM 1 to two decimal places. Record this value
as V
1
. Observe the reading on DVM 2 and record the value as V
COMP1
.
N432 A
Thermistor
sensor
A
Power meter (DUT)
Oscillator
+ input
Digital voltmeter 1 (DVM 1)
– input
V
COMP
output
V
RF
output
+ input
Digital voltmeter 2 (DVM 2)
– input
N O T E
Prior to proceeding with the following steps, ensure that the N432A and the power
meter-under-test have been switched on for at least 30 minutes.
Содержание N432A
Страница 2: ......
Страница 3: ......
Страница 20: ...XX N432A User s Guide ...
Страница 22: ...XXII N432A User s Guide ...
Страница 55: ...Operating Information 2 N432A User s Guide 33 4 Press to exit the channel setup screen ...
Страница 84: ...2 Operating Information 62 N432A User s Guide Figure 2 29 Min Hold indicated in full in the expanded window ...
Страница 90: ...2 Operating Information 68 N432A User s Guide Figure 2 36 Recall confirmation pop up dialog ...
Страница 118: ...2 Operating Information 96 N432A User s Guide ...
Страница 140: ...3 Maintenance 118 N432A User s Guide ...
Страница 152: ...4 Specifications and Characteristics 130 N432A User s Guide ...
Страница 153: ......