A
Caring for Connectors
80
Keysight N4000/1/2A Operating and Service Manual
Principles of Microwave Connector Care
Table A-4
Principles of microwave connector care
Handling and storage
DO
DO NOT
– Keep connectors clean.
– Touch mating plane surfaces.
– Extend sleeve or connector nut.
– Set connectors contact-end down.
– Use plastic end caps during storage.
Visual inspection
DO
DO NOT
– Inspect each connector carefully before every
connection.
– Use a damaged connector - EVER.
– Look for metal particles, scratches, and dents.
Cleaning
DO
DO NOT
– Try compressed air first.
– Use any abrasives.
– Clean connector threads.
– Get liquid onto plastic support beads.
Gaging
DO
DO NOT
– Clean and zero the gage before using.
– Use an out-of-spec connector.
– Use correct gage type.
– Use correct end of calibration block.
– Gage all connectors before first use.
Making connections
DO
DO NOT
– Align connectors carefully.
– Apply bending force to connection.
– Make preliminary connection lightly.
– Overtighten preliminary connection.
– Turn connector nut only to tighten.
– Twist or screw in connectors.
– Use a torque wrench for final connection.
– Tighten past “break” point of torque wrench.
Содержание N4000A SNS Series
Страница 1: ...Keysight N4000A N4001A and N4002A SNS Series Smart Noise Sources Operating and Service Manual...
Страница 10: ...10 Keysight N4000 1 2A Operating and Service Manual THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK...
Страница 12: ...12 Keysight N4000 1 2A Operating and Service Manual THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK...
Страница 14: ...14 Keysight N4000 1 2A Operating and Service Manual THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK...
Страница 44: ...4 Operating Guide 44 Keysight N4000 1 2A Operating and Service Manual Figure 4 12 Illustrated parts breakdown...