20
N2797A User’s Guide
1
Using the Probe
Dynamic Range and Offset Voltage Limitations
With the 16V
pp
(±8V) of dynamic range and ±12V of offset range, the probes have a wide
measuring range of ±20V, which allows them to be used for a wide variety of applications.
For example, if you wanted to measure small AC noise riding on top of a +10V DC signal,
you could use DC offset to position the signal at the center of the screen to keep the probe
within its dynamic range. You could then zoom in to see the detail at a higher resolution.
Trying to measure a signal out of the dynamic range may result in a clipped or distorted
waveform.
Figure 10
Dynamic Range and Offset Limits