62
N2750/1/2A User’s Guide
6
Performance Verification
Test 1. DC Input Resistance
Refer to the list of required equipment in
NOTE
Allow the probe to warm up for at least 20 minutes.
Procedure
1
Snap the socketed tip onto the probe. The socketed tip is the easiest to use in
this test.
2
Connect the probe to one of the oscilloscope’s channel inputs. This ensures
that the probe is powered.
3
Turn on the oscilloscope.
4
Using the digital multimeter, measure the DC input resistance between leads
A
and
B
. This is the differential resistance and should measure 200 k
Ω
± 2%.
5
Measure the single-ended input resistance between lead
A
and the ground
lead. The resistance should measure 100 k
Ω
± 2%.
6
Measure the single-ended input resistance between lead
B
and the ground
lead. The resistance should measure 100 k
Ω
± 2%.
7
Enter the measured values in
Содержание N2750A
Страница 1: ...Keysight N2750 1 2A Differential Probes User s Guide ...
Страница 28: ...28 N2750 1 2A User s Guide 1 Using the Probes ...
Страница 34: ...34 N2750 1 2A User s Guide 2 SPICE Models Figure 16 SPICE Model ...
Страница 38: ...38 N2750 1 2A User s Guide 2 SPICE Models ...
Страница 42: ...42 N2750 1 2A User s Guide 3 Probe Calibration ...
Страница 54: ...54 N2750 1 2A User s Guide 5 Performance Data Plots Figure 35 Frequency Response Common Mode ...
Страница 56: ...56 N2750 1 2A User s Guide 5 Performance Data Plots Figure 38 Frequency Response Common Mode ...
Страница 66: ...66 N2750 1 2A User s Guide 6 Performance Verification Figure 45 Probe Connected to Probe Adapter not to scale ...
Страница 74: ...74 N2750 1 2A User s Guide 6 Performance Verification ...