4
SCPI Programming Interface
62
M9506A-BC1 5-Slot AXIe Chassis with Integrated Controller User Guide
:DIGitizer[N]:SNUM?
Query
:DIGitizer[N]:SNUM?
Description
This query returns a string of digitizer serial number, std_string.
Example
-> dig3:snum?
<- MY12345678
:DIGitizer[N]:MODel?
Query
:DIGitizer[N]:MODel?
Description
This query returns a string of digitizer model number, std_string.
Example
-> dig3:mod?
<- M9710A
:CONFigure:MEASurement:SETTing:SAVE
Command
:CONFigure:MEASurement:SETTing:SAVE
Description
This command saves the current measurement setting into local, non-persistent
memory for easy recall. A maximum of ten measurement setting slots must keep
track of measurement settings outside of digitizer as there is no listing feature of
what setting slots have been used. std_int32 [1-10].
Example
-> conf:meas:sett:save 1
:CONFigure:MEASurement:SETTing:RECall
Command
:CONFigure:MEASurement:SETTing:RECall
Description
This command recalls 1 of 10 measurement settings that has been saved
previously using “:CONF:MEAS:SETT:SAVE” command, std_int32 [1-10].
Example
-> conf:meas:sett:rec 1
NOTE
If there is a need to go back to default settings, save default config
in one of these slots. If the default config is not saved, the module
must be reset to return to the default settings if not saved.