M9506A-BC1 5-Slot AXIe Chassis with Integrated Controller User Guide
55
SCPI Programming Interface
4
The digitizer will return from command execution even if the feature is not
supported. The system error buffer returns the errors.
Query
:DIGitizer[N]:SRATe?
Description
This query returns the current value of the sample rate for digitizer as std_float64.
Example
-> dig3:srat?
<- +5.00009
:DIGitizer[N]:RSIZe[?]
Command
:DIGitizer[N]:RSIZe
Description
This command sets the record size for data acquisition per trigger event to some
integer value of 1 to maximum memory depth enabled on the digitizer.
The value can be entered as std_int32 or std_float32. The digitizer will return from
command execution even if the feature is not supported. The system error buffer
returns the errors.
Example
-> dig3:rsiz 1024
Query
:DIGitizer[N]:RSIZe?
Description
This query returns the record size for data acquisition as std_int32.
Example
-> dig3:rsiz?
<- +1024
:DIGitizer[N]:RNUMber[?]
Command
:DIGitizer[N]:RNUMber
Description
For multi-record acquisition support, this command sets the number of records
acquired for multi-record acquisition. The value is entered as std_int32. The
digitizer will return from command execution even if the feature is not supported.
The system error buffer returns the errors.
Example
-> dig3:rnum 1