M9506A-BC1 5-Slot AXIe Chassis with Integrated Controller User Guide
29
Chassis Installation
2
Connecting and Verifying Operation with Modules Installed
In chapter 1 of this guide, you should have verified basic power-on operation of the
AXIe chassis. Using
Keysight I/O libraries Connection Expert
utility for Windows or
Linux, reverify communication with chassis and installed modules following the
process outlined below. Alternatively, you may also use a user custom I/O
interface to verify operation. When connecting to the chassis the first time, make
sure that your computer and the M9506A are in the same subnet and that there is
a DHCP server on the network. Otherwise, the
Keysight Connection Expert
will be
unable to detect your device. After you have established the initial connection, you
can change the Ethernet settings to use a static IP, if desired.
Preset Ethernet settings are:
- Activated DHCP
- Activated mDNS
- Deactivated AutoIP
1
Connect Ethernet connection
2
Power up the chassis and wait for the solid green status light on the ESM.
3
Wait for the module’s Status LEDs to turn green.
4
Establish chassis communication with
Keysight Connection Expert
.
5
Keysight Connection Expert should recognize the chassis and all installed
M9703B, M9709A, and M9710A modules, and they should be listed on the
chassis Web Interface Module Configuration Page.
The chassis Shelf Manager (on the ESM) knows whether the power supply is
connected to a Low Line or a High Line AC source and the power the installed AXIe
modules are requesting. At power-up, if the modules are requesting more power
than is available, then the Shelf Manager prevents one or more modules from
powering up.
Use the chassis Web Interface (see
M9506A User Guide
, Chapter 3, “Navigating
the Web Interface”) to determine if all modules are powered on.
NOTE
The LAN connection to the ESM provides communication to the
chassis and the integrated controller - chassis, integrated
controller and digitizer firmware updates, web interface to
monitor/control the chassis and integrated controller, etc. The
integrated controller provides communication with any installed
M9703B, M9709A, and M9710A module.