Step 5 - Verify Operation of the M9391A VSA and M9381A VSG
Step 5 - Verify Operation of the M9381A VSG and M9391A VSA
Use the following procedure to verify operation of each M9381A VSG and M9391A
VSA configuration.
P
ower on the Chassis, if you haven't already.
The first step in is to conduct a Self Test on each M9381A and M9391A.
For each M9381A, open the M9381A SFP by selecting
Start > All
.
Programs > Keysight > M938x > M9381 SFP
For each M9391A, open the M9391A SFP by selecting
Start > All
.
Programs > Keysight > M9391 > M9391 SFP
For each SFP, you are presented with the "Connect to Instrument"
dialog. Use
to select all of the modules that are
Ctrl/Select
components of the M9381A or M9391A, then press
. Leave the
Connect
SFPs open. For example:
M9381A
M9391A