
54
Keysight M8040A High-Performance BERT Getting Started Guide
2
Basic Setup for M8040A
Hardware Connections
Make the hardware connections as described below:
• Connect the M8057A to M8045A channel 1 remote head control.
• Connect the M8057A to M8045A channel 1 remote head controls
(P and N).
• Ensure that the chassis is NOT powered up or connected to a power
source while making connections to M8057A.
• Terminate unused ports (TRG Out, Data Out and Clk Out) with 50 Ohm
to GND.