Keysight M8070A Programming Guide
329
SCPI Command Reference
5
:PLUGin:OTIMing:FETCh[:JITTer][:TOTal]:MEAN?
Syntax
:PLUGin:OTIMing:FETCh[:JITTer][:TOTal]:MEAN? ‘identifier’[,‘Location’]
Input
Parameters
‘identifier’: Specify the measurement name.
[‘Location’]: Specify location (optional).
Return Range
<NR3>
Description
This command returns the mean value for total jitter associated with a
specific measurement name identifier. This value is calculated as the
weighted average of the left edge jitter histogram. This command also
associates this parameter with a specific measurement name identifier.
This SCPI is applicable for M8041A, M8051A, M8062A and M8046A.
A measurement run on a location identifier (for example, ‘M1.DataIn1’) will
return following result.
Example
:PLUG:OTIM:FETC:MEAN? ‘MyOutputTiming’,‘M1.DataIn1’
:PLUGin:OTIMing:FETCh[:JITTer][:TOTal]:PTPeak?
Syntax
:PLUGin:OTIMing:FETCh[:JITTer][:TOTal]:PTPeak? ‘identifier’[,‘Location’]
Input
Parameters
‘identifier’: Specify the measurement name.
[‘Location’]: Specify location (optional).
Return Range
<NR3>
Description
This command returns the peak to peak value for total jitter associated
with a specific measurement name identifier. This value is calculated as
the pulse period (unit interval) minus the phase margin.
This SCPI is applicable for M8041A, M8051A, M8062A and M8046A.
A measurement run on a location identifier (for example, ‘M1.DataIn1’) will
return following result.
Example
:PLUG:OTIM:FETC:PTP? ‘MyOutputTiming’,‘M1.DataIn1’