22
6000L Series Oscilloscopes Service Guide
2
Testing Performance
To construct the test connector (for use with MSO models only)
Keysight 6000L Series Oscilloscopes that have digital channels enabled require
the test connector described below. Follow the steps to build the test connector.
1
Obtain a BNC connector and an 8-by-2 section of Berg strip.
2
On one side of the Berg strip, solder a jumper wire to all of the pins (shown in
3
On the other side of the Berg strip, solder another jumper wire to all of the pins.
4
Solder the center of the BNC connector to a center pin on one of the rows on
the Berg strip.
5
Solder the ground tab of the BNC connector to a center pin on the other row on
the Berg strip.
Table 3
Materials required to construct the test connectors
Description
Recommended Part
Qty
BNC (f) Connector
Keysight 1250-1032 or
Pomona 4578
1
Berg Strip, 8-by-2
1
Jumper wire