:MEASure Commands
21
Keysight InfiniiVision M9241/42/43A PXIe Oscilloscopes SCPI Programmer's Guide
497
:MEASure:PWIDth
Command Syntax
:MEASure:PWIDth [<source>]
<source> ::= {CHANnel<n> | FUNCtion<m> | MATH<m> | WMEMory<r>}
<n> ::= 1 to (# of analog channels) in NR1 format
<m> ::= 1 to (# math functions) in NR1 format
<r> ::= 1 to (# ref waveforms) in NR1 format
The :MEASure:PWIDth command installs a screen measurement and starts the
positive pulse width measurement. If the optional source parameter is specified,
the current source is modified.
Query Syntax
:MEASure:PWIDth? [<source>]
The :MEASure:PWIDth? query measures and outputs the width of the displayed
positive pulse closest to the trigger reference. Pulse width is measured at the
midpoint of the upper and lower thresholds.
IF the edge on the screen closest to the trigger is falling:
THEN width = (time at trailing falling edge - time at leading rising edge)
ELSE width = (time at leading falling edge - time at leading rising edge)
Return Format
<value><NL>
<value> ::= width of positive pulse in seconds in NR3 format
See Also
•
"Introduction to :MEASure Commands"
•
•
•
NOTE
This command is not available if the source is FFT (Fast Fourier Transform).
Содержание InfiniiVision M9241A
Страница 1: ...Keysight InfiniiVision M9241 42 43A PXIe Oscilloscopes SCPI Programmer s Guide ...
Страница 52: ...52 Keysight InfiniiVision M9241 42 43A PXIe Oscilloscopes SCPI Programmer s Guide 2 Setting Up ...
Страница 178: ...178 Keysight InfiniiVision M9241 42 43A PXIe Oscilloscopes SCPI Programmer s Guide 5 Commands Quick Reference ...
Страница 246: ...246 Keysight InfiniiVision M9241 42 43A PXIe Oscilloscopes SCPI Programmer s Guide 7 Root Commands ...
Страница 268: ...268 Keysight InfiniiVision M9241 42 43A PXIe Oscilloscopes SCPI Programmer s Guide 8 ACQuire Commands ...
Страница 280: ...280 Keysight InfiniiVision M9241 42 43A PXIe Oscilloscopes SCPI Programmer s Guide 9 CALibrate Commands ...
Страница 354: ...354 Keysight InfiniiVision M9241 42 43A PXIe Oscilloscopes SCPI Programmer s Guide 15 EXTernal Trigger Commands ...
Страница 412: ...412 Keysight InfiniiVision M9241 42 43A PXIe Oscilloscopes SCPI Programmer s Guide 17 FUNCtion m Commands ...
Страница 416: ...416 Keysight InfiniiVision M9241 42 43A PXIe Oscilloscopes SCPI Programmer s Guide 18 HCOPY Commands ...
Страница 442: ...442 Keysight InfiniiVision M9241 42 43A PXIe Oscilloscopes SCPI Programmer s Guide 20 MARKer Commands ...
Страница 588: ...588 Keysight InfiniiVision M9241 42 43A PXIe Oscilloscopes SCPI Programmer s Guide 23 MTESt Commands ...
Страница 692: ...692 Keysight InfiniiVision M9241 42 43A PXIe Oscilloscopes SCPI Programmer s Guide 24 POWer Commands ...
Страница 1062: ...1062 Keysight InfiniiVision M9241 42 43A PXIe Oscilloscopes SCPI Programmer s Guide 30 TIMebase Commands ...
Страница 1250: ...1250 Keysight InfiniiVision M9241 42 43A PXIe Oscilloscopes SCPI Programmer s Guide 33 WGEN w Commands ...
Страница 1306: ...1306 Keysight InfiniiVision M9241 42 43A PXIe Oscilloscopes SCPI Programmer s Guide 35 Obsolete and Discontinued Commands ...
Страница 1314: ...1314 Keysight InfiniiVision M9241 42 43A PXIe Oscilloscopes SCPI Programmer s Guide 36 Error Messages ...
Страница 1376: ...1376 Keysight InfiniiVision M9241 42 43A PXIe Oscilloscopes SCPI Programmer s Guide 39 More About Oscilloscope Commands ...