What's New
1
Keysight InfiniiVision M9241/42/43A PXIe Oscilloscopes SCPI Programmer's Guide
43
Command Differences From 3000T X-Series Oscilloscopes
The Keysight InfiniiVision M9241/42/43A PXIe oscilloscopes command set is most
closely related to the InfiniiVision 3000T X-Series oscilloscopes (and the
4000 X-Series, 3000 X-Series, 7000A/B Series, 6000 Series, and
54620/54640 Series oscilloscopes before them).
The main differences between the version 7.00 programming command set for the
InfiniiVision M9241/42/43A PXIe oscilloscopes and the 4.08 programming
command set for the InfiniiVision 3000T X-Series oscilloscopes are related to:
• Two analog channels in a module (3000T X-Series oscilloscopes can have two
or four analog channels).
• No digital channels (or pods, buses, or BTIMing or BSTate math functions).
• No demo signals.
• Adds 10 MHz REF connector (similar to 4000 X-Series oscilloscopes).
• Adds precision measurements and math functions feature.
• Adds commands for setting up coordinated triggers between multiple PXIe
oscilloscope modules in a chassis.
• No USB triggering.
• No XY or Roll timebase modes.
• Transparent backgrounds for dialog boxes are not available.
• No dedicated FFT function, but you can choose FFT math function operations.
• Adds FFT Phase math function (to previous FFT Magnitude math function).
• FFT detectors are added.
• FFT measurements are added: Adjacent Channel Power Ratio (ACPR), Channel
Power, Occupied Bandwidth, and Total Harmonic Distortion.
• No LINE source for edge trigger.
• No external input on pattern trigger (just the two analog channels).
• Single-lane decoding supported for IIC and UART/RS-232.
• No support for I2S, SPI, or FlexRay serial decode and triggering.
• Internal file system locations are now on the PXIe chassis controller PC (instead
of /Agilent Flash/).
• Frequency Response Analysis feature is added.
• No printer setup (:HARDcopy comand) from within the remote interface.
• No EXTernal source allowed for serial bus sources (as compared to DSO models
of the 3000T X-Series).
• System dates and times are set by the PXIe chassis controller PC.
Содержание InfiniiVision M9241A
Страница 1: ...Keysight InfiniiVision M9241 42 43A PXIe Oscilloscopes SCPI Programmer s Guide ...
Страница 52: ...52 Keysight InfiniiVision M9241 42 43A PXIe Oscilloscopes SCPI Programmer s Guide 2 Setting Up ...
Страница 178: ...178 Keysight InfiniiVision M9241 42 43A PXIe Oscilloscopes SCPI Programmer s Guide 5 Commands Quick Reference ...
Страница 246: ...246 Keysight InfiniiVision M9241 42 43A PXIe Oscilloscopes SCPI Programmer s Guide 7 Root Commands ...
Страница 268: ...268 Keysight InfiniiVision M9241 42 43A PXIe Oscilloscopes SCPI Programmer s Guide 8 ACQuire Commands ...
Страница 280: ...280 Keysight InfiniiVision M9241 42 43A PXIe Oscilloscopes SCPI Programmer s Guide 9 CALibrate Commands ...
Страница 354: ...354 Keysight InfiniiVision M9241 42 43A PXIe Oscilloscopes SCPI Programmer s Guide 15 EXTernal Trigger Commands ...
Страница 412: ...412 Keysight InfiniiVision M9241 42 43A PXIe Oscilloscopes SCPI Programmer s Guide 17 FUNCtion m Commands ...
Страница 416: ...416 Keysight InfiniiVision M9241 42 43A PXIe Oscilloscopes SCPI Programmer s Guide 18 HCOPY Commands ...
Страница 442: ...442 Keysight InfiniiVision M9241 42 43A PXIe Oscilloscopes SCPI Programmer s Guide 20 MARKer Commands ...
Страница 588: ...588 Keysight InfiniiVision M9241 42 43A PXIe Oscilloscopes SCPI Programmer s Guide 23 MTESt Commands ...
Страница 692: ...692 Keysight InfiniiVision M9241 42 43A PXIe Oscilloscopes SCPI Programmer s Guide 24 POWer Commands ...
Страница 1062: ...1062 Keysight InfiniiVision M9241 42 43A PXIe Oscilloscopes SCPI Programmer s Guide 30 TIMebase Commands ...
Страница 1250: ...1250 Keysight InfiniiVision M9241 42 43A PXIe Oscilloscopes SCPI Programmer s Guide 33 WGEN w Commands ...
Страница 1306: ...1306 Keysight InfiniiVision M9241 42 43A PXIe Oscilloscopes SCPI Programmer s Guide 35 Obsolete and Discontinued Commands ...
Страница 1314: ...1314 Keysight InfiniiVision M9241 42 43A PXIe Oscilloscopes SCPI Programmer s Guide 36 Error Messages ...
Страница 1376: ...1376 Keysight InfiniiVision M9241 42 43A PXIe Oscilloscopes SCPI Programmer s Guide 39 More About Oscilloscope Commands ...