:ACQuire Commands
7
Keysight InfiniiVision 3000T X-Series Oscilloscopes Programmer's Guide
243
:ACQuire:TYPE
Command Syntax
:ACQuire:TYPE <type>
<type> ::= {NORMal | AVERage | HRESolution | PEAK}
The :ACQuire:TYPE command selects the type of data acquisition that is to take
place. The acquisition types are:
• NORMal — sets the oscilloscope in the normal mode.
• AVERage — sets the oscilloscope in the averaging mode. You can set the count
by sending the :ACQuire:COUNt command followed by the number of averages.
In this mode, the value for averages is an integer from 1 to 65536. The COUNt
value determines the number of averages that must be acquired.
The AVERage type is not available when in segmented memory mode
(:ACQuire:MODE SEGMented).
• HRESolution — sets the oscilloscope in the high-resolution mode (also known
as
smoothing
). This mode is used to reduce noise at slower sweep speeds
where the digitizer samples faster than needed to fill memory for the displayed
time range.
For example, if the digitizer samples at 200 MSa/s, but the effective sample
rate is 1 MSa/s (because of a slower sweep speed), only 1 out of every 200
samples needs to be stored. Instead of storing one sample (and throwing others
away), the 200 samples are averaged together to provide the value for one
display point. The slower the sweep speed, the greater the number of samples
that are averaged together for each display point.
• PEAK — sets the oscilloscope in the peak detect mode. In this mode,
:ACQuire:COUNt has no meaning.
The AVERage and HRESolution types can give you extra bits of vertical resolution.
See the
User's Guide
for an explanation. When getting waveform data acquired
using the AVERage and HRESolution types, be sure to use the WORD or ASCii
waveform data formats to get the extra bits of vertical resolution.
Query Syntax
:ACQuire:TYPE?
The :ACQuire:TYPE? query returns the current acquisition type.
Return Format
<acq_type><NL>
<acq_type> ::= {NORM | AVER | HRES | PEAK}
See Also
•
"Introduction to :ACQuire Commands"
NOTE
The obsolete command ACQuire:TYPE:REALtime is functionally equivalent to sending
ACQuire:MODE RTIME; TYPE NORMal.
Содержание InfiniiVision 3000T X Series
Страница 1: ...Keysight InfiniiVision 3000T X Series Oscilloscopes Programmer s Guide...
Страница 40: ...40 Keysight InfiniiVision 3000T X Series Oscilloscopes Programmer s Guide 1 What s New...
Страница 50: ...50 Keysight InfiniiVision 3000T X Series Oscilloscopes Programmer s Guide 2 Setting Up...
Страница 64: ...64 Keysight InfiniiVision 3000T X Series Oscilloscopes Programmer s Guide 3 Getting Started...
Страница 164: ...164 Keysight InfiniiVision 3000T X Series Oscilloscopes Programmer s Guide 4 Commands Quick Reference...
Страница 254: ...254 Keysight InfiniiVision 3000T X Series Oscilloscopes Programmer s Guide 8 BUS n Commands...
Страница 286: ...286 Keysight InfiniiVision 3000T X Series Oscilloscopes Programmer s Guide 10 CHANnel n Commands...
Страница 306: ...306 Keysight InfiniiVision 3000T X Series Oscilloscopes Programmer s Guide 12 DEMO Commands...
Страница 314: ...314 Keysight InfiniiVision 3000T X Series Oscilloscopes Programmer s Guide 13 DIGital d Commands...
Страница 344: ...344 Keysight InfiniiVision 3000T X Series Oscilloscopes Programmer s Guide 16 EXTernal Trigger Commands...
Страница 352: ...352 Keysight InfiniiVision 3000T X Series Oscilloscopes Programmer s Guide 17 FFT Commands FFT WINDow on page 362...
Страница 364: ...364 Keysight InfiniiVision 3000T X Series Oscilloscopes Programmer s Guide 17 FFT Commands...
Страница 402: ...402 Keysight InfiniiVision 3000T X Series Oscilloscopes Programmer s Guide 18 FUNCtion m Commands...
Страница 420: ...420 Keysight InfiniiVision 3000T X Series Oscilloscopes Programmer s Guide 19 HARDcopy Commands...
Страница 442: ...442 Keysight InfiniiVision 3000T X Series Oscilloscopes Programmer s Guide 21 MARKer Commands...
Страница 466: ...466 Keysight InfiniiVision 3000T X Series Oscilloscopes Programmer s Guide 22 MEASure Commands MEASure CLEar on page 464...
Страница 524: ...524 Keysight InfiniiVision 3000T X Series Oscilloscopes Programmer s Guide 22 MEASure Commands...
Страница 548: ...548 Keysight InfiniiVision 3000T X Series Oscilloscopes Programmer s Guide 23 MEASure Power Commands...
Страница 656: ...656 Keysight InfiniiVision 3000T X Series Oscilloscopes Programmer s Guide 26 POWer Commands...
Страница 666: ...666 Keysight InfiniiVision 3000T X Series Oscilloscopes Programmer s Guide 27 RECall Commands...
Страница 888: ...888 Keysight InfiniiVision 3000T X Series Oscilloscopes Programmer s Guide 29 SBUS n Commands...
Страница 978: ...978 Keysight InfiniiVision 3000T X Series Oscilloscopes Programmer s Guide 30 SEARch Commands...
Страница 1006: ...1006 Keysight InfiniiVision 3000T X Series Oscilloscopes Programmer s Guide 32 TIMebase Commands...
Страница 1232: ...1232 Keysight InfiniiVision 3000T X Series Oscilloscopes Programmer s Guide 37 Obsolete and Discontinued Commands...
Страница 1240: ...1240 Keysight InfiniiVision 3000T X Series Oscilloscopes Programmer s Guide 38 Error Messages...
Страница 1264: ...1264 Keysight InfiniiVision 3000T X Series Oscilloscopes Programmer s Guide 39 Status Reporting...
Страница 1274: ...1274 Keysight InfiniiVision 3000T X Series Oscilloscopes Programmer s Guide 40 Synchronizing Acquisitions...
Страница 1404: ...1404 Keysight InfiniiVision 3000T X Series Oscilloscopes Programmer s Guide 42 Programming Examples...