178
Keysight InfiniiVision 3000 X-Series Oscilloscopes User's Guide
11
Trigger Mode/Coupling
2
In the Trigger Mode and Coupling Menu, press the
Coupling
softkey; then, turn
the Entry knob to select:
•
DC
coupling — allows DC and AC signals into the trigger path.
•
AC
coupling — places a 10 Hz high-pass filter in the trigger path removing
any DC offset voltage from the trigger waveform.
The high-pass filter in the External Trigger input path is 50 Hz for all models.
Use AC coupling to get a stable edge trigger when your waveform has a
large DC offset.
•
LF
(low frequency)
Reject
coupling — adds a high-pass filter with the 3-dB
point at 50 kHz in series with the trigger waveform.
Low frequency reject removes any unwanted low frequency components
from a trigger waveform, such as power line frequencies, etc., that can
interfere with proper triggering.
Use
LF Reject
coupling to get a stable edge trigger when your waveform has
low frequency noise.
•
TV
coupling — is normally grayed-out, but is automatically selected when TV
trigger is enabled in the Trigger Menu.
Note that Trigger Coupling is independent of Channel Coupling (see
Pass
Band
50 kHz
0 dB
3 dB down point
DC
Содержание InfiniiVision 3000 X Series
Страница 1: ...Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide...
Страница 8: ...8 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide...
Страница 24: ...24 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide...
Страница 48: ...48 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 1 Getting Started...
Страница 62: ...62 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 2 Horizontal Controls...
Страница 72: ...72 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 3 Vertical Controls...
Страница 122: ...122 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 6 Digital Channels...
Страница 128: ...128 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 7 Serial Decode...
Страница 134: ...134 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 8 Display Settings...
Страница 182: ...182 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 11 Trigger Mode Coupling...
Страница 200: ...200 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 12 Acquisition Control...
Страница 210: ...210 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 13 Cursors...
Страница 252: ...252 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 15 Mask Testing...
Страница 256: ...256 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 16 Digital Voltmeter...
Страница 290: ...290 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 19 Print Screens...
Страница 324: ...324 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 21 Web Interface...
Страница 344: ...344 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 22 Reference...
Страница 370: ...370 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 24 FlexRay Triggering and Serial Decode...
Страница 424: ...424 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 28 UART RS232 Triggering and Serial Decode...