Troubleshooting
4
90000 A-Series Service Guide
77
Checking Probe Power Outputs
Probe power outputs are on the front panel, under each BNC input.
Refer to the following figure to check the power output at the connectors. Measure
the +3 V, -3 V, +12 V, and -12 V voltages with respect to the ground terminal on
the front panel, located near the Aux Out BNC.
Do not try to measure voltages at pins 3 through 7.
Any failure is likely caused by a problem with the probe power and control
assembly, the AutoProbe flex cable, the probe power and control cable, or the
power board.
Pin
Supply
1
+3V
2
–3V
3
Offset
4
Data
5 & ring
Probe ID
6
Clk
7
R
p
8
–12 V
9
+12 V
Содержание Infiniium 90000A Series
Страница 1: ...Keysight Infiniium 90000 A Series Oscilloscopes Service Guide...
Страница 6: ...6 90000 A Series Oscilloscopes Service Guide...
Страница 12: ...12 90000 A Series Service Guide 1 General Information...
Страница 18: ...18 90000 A Series Service Guide 2 Calibration...
Страница 110: ...110 90000 A Series Service Guide 5 Replacing Assemblies P owerharness between backplaneand m otherboard O n O ff ca ble...
Страница 116: ...116 90000 A Series Service Guide 5 Replacing Assemblies...
Страница 119: ...Replaceable Parts 6 90000 A Series Service Guide 119 Exploded Views Front Frame and Front Panel...
Страница 120: ...120 90000 A Series Service Guide 6 Replaceable Parts Fan and Acquisition Assembly...
Страница 121: ...Replaceable Parts 6 90000 A Series Service Guide 121 Power Supply and PC Motherboard...
Страница 122: ...122 90000 A Series Service Guide 6 Replaceable Parts Sleeve and Accessory Pouch...
Страница 128: ...128 90000 A Series Service Guide 7 Theory of Operation Oscilloscope block diagram...
Страница 135: ......