Keysight Technologies EDU33210 Series Скачать руководство пользователя страница 5

U.S. Government Rights

The Software is “commercial computer software,” as defined by Federal Acquisition Regulation (“FAR”) 2.101.
Pursuant to FAR 12.212 and 27.405-3 and Department of Defense FAR Supplement (“DFARS”) 227.7202, the U.S.
government acquires commercial computer software under the same terms by which the software is customarily
provided to the public. Accordingly, Keysight provides the Software to U.S. government customers under its
standard commercial license, which is embodied in its End User License Agreement (EULA), a copy of which can be
found at

http://www.keysight.com/find/sweula

The license set forth in the EULA represents the exclusive authority

by which the U.S. government may use, modify, distribute, or disclose the Software. The EULA and the license set
forth therein, does not require or permit, among other things, that Keysight: (1) Furnish technical information related
to commercial computer software or commercial computer software documentation that is not customarily provided
to the public; or (2) Relinquish to, or otherwise provide, the government rights in excess of these rights customarily
provided to the public to use, modify, reproduce, release, perform, display, or disclose commercial computer
software or commercial computer software documentation. No additional government requirements beyond those
set forth in the EULA shall apply, except to the extent that those terms, rights, or licenses are explicitly required from
all providers of commercial computer software pursuant to the FAR and the DFARS and are set forth specifically in
writing elsewhere in the EULA. Keysight shall be under no obligation to update, revise or otherwise modify the
Software. With respect to any technical data as defined by FAR 2.101, pursuant to FAR 12.211 and 27.404.2 and
DFARS 227.7102, the U.S. government acquires no greater than Limited Rights as defined in FAR 27.401 or DFAR
227.7103-5 (c), as applicable in any technical data.

Third Party Licenses

Portions of this software are licensed by third parties including open source terms and conditions. For more
information, please contact Keysight support,

www.keysight.com/find/assist

.

Waste Electrical and Electronic Equipment (WEEE)

This product complies with the WEEE Directive) marketing requirement. The affixed product label (see below)
indicates that you must not discard this electrical/electronic product in domestic household waste.

Product Category: With reference to the equipment types in the WEEE directive Annex 1, this product is classified as
“Monitoring and Control instrumentation” product. Do not dispose in domestic household waste.

To return unwanted products, contact your local Keysight office, or see

about.keysight.com/en/companyinfo/environment/takeback.shtml

for more information.

Declarations of Conformity

Declarations of Conformity for this product and for other Keysight products may be downloaded from the Web. Go to

https://regulations.about.keysight.com/DoC/default.htm

. You can then search by product number to find the latest

Declaration of Conformity.

Keysight EDU33210 Series Service Guide

5

Содержание EDU33210 Series

Страница 1: ...Service Guide EDU33210 Series Trueform Arbitrary Waveform Generators...

Страница 2: ...Installing the Knob 13 2 Performance Verification 14 Performance Verification 15 Recommended Test Equipment 16 Test Considerations 16 Performance Verification Tests 17 Internal timebase verification...

Страница 3: ...ude high impedance adjustment 35 8 dB range flatness adjustment 37 24 dB range flatness adjustment 39 Channel 2 Adjustments EDU33212A only 41 Self calibration adjustment channel 2 41 Output impedance...

Страница 4: ...Y APPLICABLE LAW KEYSIGHT DISCLAIMS ALL WARRANTIES EITHER EXPRESS OR IMPLIED WITH REGARD TO THIS MANUAL AND ANY INFORMATION CONTAINED HEREIN INCLUDING BUT NOT LIMITED TO THE IMPLIED WARRANTIES OF MERC...

Страница 5: ...ed from all providers of commercial computer software pursuant to the FAR and the DFARS and are set forth specifically in writing elsewhere in the EULA Keysight shall be under no obligation to update...

Страница 6: ...Do not proceed beyond a CAUTION notice until the indicated conditions are fully understood and met A WARNING notice denotes a hazard It calls attention to an operating procedure practice or the like...

Страница 7: ...laceable Parts Battery Replacement Removing Installing the Knob This chapter provides the specifications and service information on cleaning troubleshooting repair and replaceable parts of the EDU3321...

Страница 8: ...ey will arrange to have your unit repaired or replaced and can provide warranty or repair cost information where applicable Ask the Keysight Technologies Service Center for shipping instructions inclu...

Страница 9: ...during service operations Disassemble instruments only in a static free work area Use a conductive work area to reduce static charges Use a conductive wrist strap to reduce static charge accumulation...

Страница 10: ...r messages include a failing channel number 1 or 2 shown as n in the messages below Error Code Error Messages 601 Self test failed real time clock settings lost 602 Self test failed main CPU power sup...

Страница 11: ...ed service trained personnel who are aware of the hazards involved should remove instrument cov ers Always disconnect the power cable and any external circuits before removing the instrument cover Som...

Страница 12: ...river and carefully pry up on the side of the battery 6 Install the new battery Make sure that the positive side is facing up Place the battery under the small spring clips see red circle below then p...

Страница 13: ...ions Visual 1 Pull out the knob to remove the knob from the front panel 2 To install back the knob push the knob back in the shaft NOTE Make sure to follow the shaft orientation before pushing the kno...

Страница 14: ...quipment Test Considerations Performance Verification Tests This chapter contains the performance verification procedures which verify that the EDU33210 Series waveform generator is operating within i...

Страница 15: ...ion tests Self test A series of internal verification tests that give a high confidence that the instrument is operational Performance verification tests An extensive set of tests that are recommended...

Страница 16: ...0 termination Keysight part number 0960 0301 Adapter Coaxial BNC T connector Cable Dual banana m to dual banana m Cable RG58 coaxial cable with BNC m to dual banana adapter Cable RG58 coaxial cable wi...

Страница 17: ...0 of the specifications as the verification limits Ensure that you have read Test Considerations before running the performance verification tests Internal timebase verification These tests verify out...

Страница 18: ...s 1 kHz 400 mVrms 0 008354 Vrms High Z Sine 400 mVrms 1 kHz 400 mVrms 0 008354 Vrms High Z Sine 1 Vrms 1 kHz 1 Vrms 0 020354 Vrms High Z Sine 2 5 Vrms 1 kHz 2 5 Vrms 0 050354 Vrms High Z Sine 7 Vrms 1...

Страница 19: ...r Be sure the output impedance is set to High Z and the output is enabled Waveform Generator Measurement Output Setup Function Voltage Nominal Error a High Z DC 0 0 V 0 0 VDC 0 005 VDC High Z DC 500 m...

Страница 20: ...e flatness reference point at 1 kHz 4 Convert multimeter error to ratio using equation below This value is required for next measurement point Reference point accuracy 1 kHz dB 20 x log multimeter mea...

Страница 21: ...asure the amplitude by using power meter and power sensor Waveform Generator Measurement Output Load Function Amplitude Frequency Nominal Error 50 Sine 1 200 Vrms 500 0 kHz 0 dB 0 15 dB 1 000 MHz 2 00...

Страница 22: ...he flatness reference point at 1 kHz 4 Convert multimeter error to ratio using equation below This value is required for next measurement point Reference point accuracy 1 kHz dB 20 x log multimeter me...

Страница 23: ...measure the amplitude by using power meter and power sensor Waveform Generator Measurement Output Load Function Amplitude Frequency Nominal Error 50 Sine 0 190 Vrms 500 0 kHz 0 dB 0 15 dB 1 000 MHz 2...

Страница 24: ...24 Keysight EDU33210 Series Service Guide...

Страница 25: ...Process Calibration Security Calibration Count Calibration Message Calibration Procedure Channel 1 Adjustments Channel 2 Adjustments This chapter contains information on adjustments performed after a...

Страница 26: ...service at competitive prices Calibration interval The recommended calibration interval for Keysight EDU33210 Series waveform generator is one year Time required for calibration For incoming instrumen...

Страница 27: ...ty code and calibration count in the instrument s maintenance records 7 Perform the performance verification tests to verify the calibration Calibration Security The instrument has a calibration passc...

Страница 28: ...increments up to a maximum of 232 1 after which it rolls over to 0 The calibration count can only be read remotely From the remote interface CAL COUN Calibration Message You can use the CALibration ST...

Страница 29: ...failed Chan n offset DAC cal with attenuator gain too low too high meas_value Self calibration failed Chan n offset DAC cal no attenuator gain too low too high meas_value Computed gain calibration fa...

Страница 30: ...ide If you abort a calibration in progress all calibration constants for the selected function range are lost If power is turned off when the instrument is attempting to write new calibration constant...

Страница 31: ...ing any additional adjustments Be sure to unlock the instrument and follow the requirements listed in Test Considerations before beginning any adjustments 1 Enter command CALibration SeTup 1 to begin...

Страница 32: ...uency for each setup in the following table Setup Frequency Amplitude SCPI command 2 10 MHz 1 Vpp Output frequency is slightly less than 10 MHz CAL SET 2 3 10 MHz Output frequency is slightly more tha...

Страница 33: ...ter as shown below 2 Set the multimeter to display 5 digits and set the function to DCV 3 Enter the following setup Nominal Setup Setup DC level SCPI command 6 a 1 0 VDC 10 Calibrates the internal ADC...

Страница 34: ...are generated with and without the post amplifier attenuator 1 Set the multimeter to measure offset compensated 4 wire Ohms Set the multimeter to use 100 NPLC integ ration Connect the Ohms Source and...

Страница 35: ...range CAL SET 11 12 0 007 V Output of 64 dB range CAL SET 12 13 a 0 007 V Output of 64 dB range CAL SET 13 14 0 017 V Output of 56 dB range CAL SET 14 15 a 0 017 V Output of 56 dB range CAL SET 15 16...

Страница 36: ...AL SET 35 36 0 68 V Output of 24 dB range CAL SET 36 37 a 0 68 V Output of 24 dB range CAL SET 37 a Constants are stored after completing this setup 3 Enter the value measured using SCPI command 4 To...

Страница 37: ...sured using SCPI command 4 For Setup 47 record down the multimeter measurement value during 100 kHz frequency output Multimeter measurement 100 kHz ______________Vrms 5 Without resetting and setting t...

Страница 38: ...red power level from dBm to Vrms 12 Enter the value measured calibration value using SCPI command Calibration Value Power meter converted value in Vrms Crossover 100 kHz in Vrms Step 9 13 To proceed a...

Страница 39: ...sured using SCPI command 4 For Setup 39 record down the multimeter measurement value during 100 kHz frequency output Multimeter measurement 100 kHz ______________Vrms 5 Without resetting and setting t...

Страница 40: ...ET 43 44 a 25 MHz CAL SET 44 45 a 30 MHz CAL SET 45 a Constants are stored after completing this setup 11 Convert the measured power level from dBm to Vrms 12 Enter the value measured calibration valu...

Страница 41: ...ng setup Setup SCPI command 54 a Self calibration The output is disabled CAL SET 54 a Constants are stored after completing this setup 2 To proceed a If your calibration procedures require you to veri...

Страница 42: ...channel output as shown below 2 Use the multimeter to make a 4 wire resistance measurement at the front panel output connector for each setup in the following table The expected measured value is app...

Страница 43: ...2 dB range CAL SET 58 59 0 007 V Output of 64 dB range CAL SET 59 60 a 0 007 V Output of 64 dB range CAL SET 60 61 0 017 V Output of 56 dB range CAL SET 61 62 a 0 017 V Output of 56 dB range CAL SET 6...

Страница 44: ...32 dB range High DC range CAL SET 82 83 0 68 V Output of 24 dB range High DC range CAL SET 83 84 a 0 68 V Output of 24 dB range High DC range CAL SET 84 a Constants are stored after completing this s...

Страница 45: ...e measured using SCPI command 4 For Setup 94 record down the multimeter measurement value during 100 kHz frequency output Multimeter measurement 100 kHz ______________Vrms 5 Without resetting and sett...

Страница 46: ...onstants are stored after completing this setup 11 Convert the measured power level from dBm to Vrms 12 Enter the value measured calibration value using SCPI command Calibration Value Power meter conv...

Страница 47: ...ue measured using SCPI command 4 For Setup 86 record down the multimeter measurement value during 100 kHz frequency output Multimeter measurement 100 kHz ______________Vrms 5 Without resetting and set...

Страница 48: ...ET 90 91 a 25 MHz CAL SET 91 92 a 30 MHz CAL SET 92 a Constants are stored after completing this setup 11 Convert the measured power level from dBm to Vrms 12 Enter the value measured calibration valu...