Keysight 

E8267D PSG Signal Generators 

Upgrade Kit for Option 601 and 602 

(Digital Bus Baseband Generator with 8 

and 64 Msample Memory, respectively) 

Kit Part Numbers E8267DK-601 and E8267DK-602

Installation Note

Part Number

:

 

E8251-90388

 

Printed in USA   A

ug

ust 

 

2014

 

Содержание E8267D

Страница 1: ...sight Technologies For more in formation go to www keysight com Keysight E8267D PSG Signal Generators Upgrade Kit for Option 601 and 602 Digital Bus Baseband Generator with 8 and 64 Msample Memory res...

Страница 2: ...this material including but not limited to the implied warranties of merchantability and fitness for a particular purpose Keysight Technologies shall not be liable for errors contained herein or for...

Страница 3: ...2 Initialize the signal generator memory 3 Remove the outer and inner signal generator covers 4 Install the A14 Digital Bus Baseband Generator board 5 Verify proper installation of Option 601 602 6 R...

Страница 4: ...Electrostatic discharge ESD can damage or destroy electronic components All work on electronic assemblies should be performed at a static safe workstation Item Quantity Description Part Number 1 1 In...

Страница 5: ...2 Run the signal generator self test by pressing Utility Instrument Info Help Mode Self Test Run Complete Self Test Upon completion a summary of the self test will be displayed Use the E8257D 67D PSG...

Страница 6: ...nter screws 3 on the four rear panel feet 4 4 Remove the four bottom feet 5 from the cover by lifting the tab and sliding the foot toward the tab 5 Place the signal generator on its side 6 Tilt the si...

Страница 7: ...7 Figure 1 Outer Cover Removal...

Страница 8: ...8 Figure 2 Inner Top Cover Removal...

Страница 9: ...Figure 3 and Figure 4 for this procedure 1 Disconnect the power cord 2 Disconnect the W66 ribbon cable from A18P2 3 Using the standard screwdriver unlock the A41 door latch screw by rotating counter c...

Страница 10: ...4 Connect the W64 ribbon cable to A17P2 5 Connect the W90 cable end with foam attached to it to A14J21 6 Connect the W90 cable to A17P1 7 Connect the W90 cable to A14J21 8 Connect the W64 ribbon cable...

Страница 11: ...11 Figure 5 Baseband Cover Removal 1 1 1 1...

Страница 12: ...12 Figure 6 A14 Internal Baseband Generator Board Figure 7 A17 Baseband Generator Interface Board A14 W65 A14J801 W64 A14J13 W90 A14J21 W64 A17P2 W90 A17P1...

Страница 13: ...N Make sure the angled side of the cable connector on W90 is facing up Failure to do so could result in damage to the A17 Rear Panel Interface Board connector Figure 8 Digital Bus Cable Installation W...

Страница 14: ...15 and T 20 screws to 21 in lbs 4 Attach the Digital Bus label over DIG I Q I O on the rear panel Run the I Q Calibration 1 Turn the instrument on and let it warm up for 30 minutes 2 Press I Q I Q Cal...