Table C-3
Status Bit Definitions of the Operation Status Register
Bit
Position
Name
Description
Condition Register
Event Register
0
Not used
Always 0
Always 0
1
Settling
Set to “1” during the waiting time
to stabilize the measurement
signal.
Set to “1” after the waiting time
needed to stabilize the
measurement signal has elapsed.
2
Ranging
Set to “1” while switching the
measurement range.
Set to “1” when measurement
range switching is completed.
3
Analog Measurement
Set to “1” during analog
measurement
*1
.
Set to “1” when analog
measurement is completed.
4
Measurement
Set to “1” during measurement
*2
.
Set to “1” when measurement is
completed.
5
Waiting for Trigger
Set to “1” while the instrument is
waiting for a trigger
*3
.
Set to “1” when the instrument
starts waiting for a trigger.
6
Not used
Always 0
Always 0
7
Correction
Set to “1” during correction data
measurement.
Set to “1” when the correction data
measurement is completed.
8
Data buffer 1
Set to “1” while data buffer 1 fills.
Set to “1” when data buffer 1 is
full.
9
Data buffer 2
Set to “1” while data buffer 2 fills.
Set to “1” when data buffer 2 is
full.
10
Data buffer 3
Set to “1” while data buffer 3 fills.
Set to “1” when data buffer 3 is
full.
11
Not used
Always 0
Always 0
12
Self-test
Set to “1” during self-test.
Set to “1” when self-test is
completed.
13 to 15
Not used
Always 0
Always 0
*1. This is when the handler interface’s /INDEX signal is active.
*2. This is when the handler interface’s /EOM signal is active.
*3. This is when the trigger system is in trigger wait state. For more information on the trigger
system, refer to
Table C-4
Status Bit Definitions of the Questionable Status Register
Bit
Position
Name
Description
Condition Register
Event Register
0 to 15
Not used
Always 0
Always 0
Содержание E4981A
Страница 2: ......
Страница 20: ...18 Chapter1 Making Effective Use of This Manual How To Use This Manual ...
Страница 60: ...58 Chapter4 Preparation for Accurate Measurement Executing Correction Using the Cable Correction Function ...
Страница 88: ...86 Chapter7 Sorting Based on Measured Result Comparator Function Sample Program ...
Страница 162: ...160 Chapter9 Measurement Applications Sample Programs Detecting Error Occurrence using SRQ HT Basic ...
Страница 376: ...374 AppendixE Error Messages Warning Messages WARNING ...
Страница 382: ......
Страница 383: ...This information is subject to change without notice Keysight Technologies 2008 2014 Edition 2 August 2014 E4981 90001 ...