260
Chapter 10
Command Reference
E4981A SCPI Commands
[:SENSe][:FIMPedance]:CREJect[:STATe]
Syntax
[:SENSe][:FIMPedance]:CREJect[:STATe] {ON|OFF|1|0}
[:SENSe][:FIMPedance]:CREJect[:STATe]?
Description
Enables/disables the Low C reject function.
When you enable the Low C reject function, if the measured value of the primary
parameter (Cp or Cs) is too small (equal to or less than the boundary value specified with
[:SENSe][:FIMPedance]:CREJect:LIMit
command), Low C is detected. For
information on the screen display, GPIB/USB/LAN output, and so on when Low C is
detected, refer to Appendix D, “Operations when Overload, No Contact or Low C is
Parameters
Query response
{1|0}<newline><^END>
Related commands
[:SENSe][:FIMPedance]:CREJect:LIMit on page 259
Equivalent key
sequence
[
Meas Setup
] - LOW C REJ - ON|OFF
Description
ON or 1
Enables the Low C reject function.
OFF or 0
(initial value)
Disables the Low C reject function.
Содержание E4981A
Страница 2: ......
Страница 20: ...18 Chapter1 Making Effective Use of This Manual How To Use This Manual ...
Страница 60: ...58 Chapter4 Preparation for Accurate Measurement Executing Correction Using the Cable Correction Function ...
Страница 88: ...86 Chapter7 Sorting Based on Measured Result Comparator Function Sample Program ...
Страница 162: ...160 Chapter9 Measurement Applications Sample Programs Detecting Error Occurrence using SRQ HT Basic ...
Страница 376: ...374 AppendixE Error Messages Warning Messages WARNING ...
Страница 382: ......
Страница 383: ...This information is subject to change without notice Keysight Technologies 2008 2014 Edition 2 August 2014 E4981 90001 ...