Chapter 10
187
Command Reference
E4981A SCPI Commands
10
. Comm
and
Refe
ren
ce
:CALCulate1:COMParator[:STATe]
Syntax
:CALCulate1:COMParator[:STATe] {ON|OFF|1|0}
:CALCulate1:COMParator[:STATe]?
Description
Turns ON/OFF the comparator function.
This setup is interlocked with the ON/OFF state of the signal output of the handler
interface.
NOTE
Changing the measurement parameter will automatically turn off this function. To avoid
this, you have to write a program so that this command is executed after the execution of
the measurement parameter setup command (
Parameters
Query response
{1|0}<newline><^END>
Related commands
:CALCulate1:FORMat on page 188
:CALCulate2:FORMat on page 191
Equivalent key
sequence
[
Meas Setup
] - LIMIT TABLE - COMP
Description
ON or 1
Turns ON the comparator function.
OFF or 0 (initial value)
Turns OFF the comparator function.
Содержание E4981A
Страница 2: ......
Страница 20: ...18 Chapter1 Making Effective Use of This Manual How To Use This Manual ...
Страница 60: ...58 Chapter4 Preparation for Accurate Measurement Executing Correction Using the Cable Correction Function ...
Страница 88: ...86 Chapter7 Sorting Based on Measured Result Comparator Function Sample Program ...
Страница 162: ...160 Chapter9 Measurement Applications Sample Programs Detecting Error Occurrence using SRQ HT Basic ...
Страница 376: ...374 AppendixE Error Messages Warning Messages WARNING ...
Страница 382: ......
Страница 383: ...This information is subject to change without notice Keysight Technologies 2008 2014 Edition 2 August 2014 E4981 90001 ...