![Keysight Technologies E36100 Series Скачать руководство пользователя страница 89](http://html1.mh-extra.com/html/keysight-technologies/e36100-series/e36100-series_operating-and-service-manual_1962203089.webp)
Test Record Forms
Keysight E36102A
Report number: ______________________
Date: _______________________________
Description
Lower limit
Result
Upper limit
Constant Voltage Tests
Voltage Programming
– Zero Voltage Output (V0)
-0.003 V
______________
0.003 V
– Maximum Voltage Ouput (Vmax)
5.994 V
______________
6.006 V
Voltage Readback
– Zero Voltage measured over interface
V0 - 0.003 V
______________
V0 + 0.003 V
– Maximum Voltage measured over interface
Vmax - 0.006 V
______________
Vmax + 0.006 V
CV Load Regulation (Vload - Vnoload)
-2.6 mV
______________
2.6 mV
CV Line Regulation (Vlowline - Vhighline)
-1.6 mV
______________
1.6 mV
CV Ripple and Noise
– peak-to-peak
—
______________
10 mV
– rms
—
______________
10 μV
Transient Response
—
______________
50 μs
Constant Current Tests
Current Programming
– Zero Current Output (I0)
-0.004 A
______________
0.004 A
– Maximum Current Ouput (Imax)
4.9925 A
______________
5.0075 A
Current Readback
– Zero Current measured over interface
I0 - 0.004 A
______________
I0 + 0.004 A
– Maximum Current measured over interface
Imax - 0.0065 A
______________
Imax + 0.0065 A
CC Load Regulation (Iload - Ishort)
-1250 μA
______________
1250 μA
CC Line Regulation (Ilowline - Ihighline)
-1250 μA
______________
1250 μA
Test Description
Instrument settings
Voltage Programming & Readback, Zero Voltage (V0)
0 V
5 A
Voltage Programming & Readback, Maximum Voltage (Vmax)
6 V
5 A