50
1000 X-Series Oscilloscopes Service Guide
4
Troubleshooting
To compensate passive probes
You should compensate your passive probes to match their characteristics to the
oscilloscope’s channels. A poorly compensated probe can introduce measurement
errors.
1
Perform the procedure
“To perform an Auto Scale on the Probe Comp
2
Press the channel key to which the probe is connected (
[1]
,
[2]
).
3
In the Channel Menu, press
Probe
.
4
In the Channel Probe Menu, press
Probe Check
; then, follow the instructions
on-screen.
If necessary, use a nonmetallic tool (supplied with the probe) to adjust the
trimmer capacitor on the probe for the flattest pulse possible.
On the some probes, the trimmer capacitor is the yellow adjustment on the
probe tip. On other probes, the trimmer capacitor is located on the probe BNC
connector.
Figure 7
Example pulses
5
Connect probes to all other oscilloscope channels (channel 2 of a 2-channel
oscilloscope).
6
Repeat the procedure for each channel.
The process of compensating the probes serves as a basic test to verify that the
oscilloscope is functional.
comp.cdr
Perfectly compensated
Over compensated
Under compensated
Содержание DSOX1102A
Страница 1: ...Keysight InfiniiVision 1000 X Series Oscilloscopes Service Guide ...
Страница 7: ...1000 X Series Oscilloscopes Service Guide 7 Contents Index ...
Страница 8: ...8 1000 X Series Oscilloscopes Service Guide Contents ...
Страница 10: ...10 1000 X Series Oscilloscopes Service Guide Figures ...
Страница 12: ...12 1000 X Series Oscilloscopes Service Guide Tables ...
Страница 52: ...52 1000 X Series Oscilloscopes Service Guide 5 Replacing Assemblies ...
Страница 54: ...54 1000 X Series Oscilloscopes Service Guide 6 Replaceable Parts ...
Страница 60: ...60 1000 X Series Oscilloscopes Service Guide Index ...