32
1000 X-Series Oscilloscopes Service Guide
2
Testing Performance
To verify trigger sensitivity
These tests verify the trigger sensitivity. In these tests, you will apply a sine wave
to the oscilloscope at two bandwidths: 10 MHz and the maximum bandwidth of
the oscilloscope. For each sine wave, you will decrease the amplitude of the signal
to a specified level and check to see if the oscilloscope still triggers.
The external trigger sensitivity is also tested in both the 1.6 V and 8 V ranges for
the 70 MHz and 100 MHz (DSOX) models.
Table 9
Equipment Required to Verify Trigger Sensitivity
Equipment
Critical Specifications
Recommended Model/Part
Signal Generator
10 MHz and 50 MHz, 70 MHz, or
100 MHz sine waves
Keysight N5181A
Power splitter
Outputs differ < 0.15 dB
Keysight 11667A
Power Meter
Keysight N1914A
Power Sensor
Keysight E9304A or N8482A
Cable
BNC, Qty 3
Keysight 10503A
Adapter
N (m) to BNC (f), Qty 3
Keysight 1250-0780
Feedthrough
50
Ω
BNC (f) to BNC (m)
Keysight 0960-0301
Содержание DSOX1102A
Страница 1: ...Keysight InfiniiVision 1000 X Series Oscilloscopes Service Guide ...
Страница 7: ...1000 X Series Oscilloscopes Service Guide 7 Contents Index ...
Страница 8: ...8 1000 X Series Oscilloscopes Service Guide Contents ...
Страница 10: ...10 1000 X Series Oscilloscopes Service Guide Figures ...
Страница 12: ...12 1000 X Series Oscilloscopes Service Guide Tables ...
Страница 52: ...52 1000 X Series Oscilloscopes Service Guide 5 Replacing Assemblies ...
Страница 54: ...54 1000 X Series Oscilloscopes Service Guide 6 Replaceable Parts ...
Страница 60: ...60 1000 X Series Oscilloscopes Service Guide Index ...