Testing Performance
2
1000 X-Series Oscilloscopes Service Guide
21
To verify DC vertical gain accuracy
This test verifies the accuracy of the analog channel DC vertical gain for each
channel.
In this test, you will measure the dc voltage output of an oscilloscope calibrator
using the oscilloscope’s
Average - Full Screen
voltage measurement and compare
the results with the multimeter reading.
1
Press
[Save/Recall] > Default/Erase > Factory Default
to recall the factory default
setup.
2
Set the probe attenuation to 1:1 on the analog channel you are testing (for
example,
[1] > Probe > Probe
; then, turn the
Entry knob to select
1.00 : 1
).
Table 3
DC Vertical Gain Accuracy Test Limits
Test Limits
Notes
±3% of full scale (>= 10 mV/div);
±4% of full scale (< 10 mV/div)
•
Full scale is defined as 8 mV on the 500 uV/div range.
•
Full scale on all other ranges is defined as 8 divisions
times the V/div setting.
Table 4
Equipment Required to Verify DC Vertical Gain Accuracy
Equipment
Critical Specifications
Recommended Model/Part
Oscilloscope Calibrator
3.5 mV to 70 Vdc,
0.1 V resolution
Fluke 5820A
Digital multimeter
Better than 0.01% accuracy
Keysight 34401A
Cable
BNC, Qty 2
Keysight 10503A
Shorting cap
BNC
Keysight 1250-0774
Adapter
BNC (f) to banana (m)
Keysight 1251-2277
Adapter
BNC tee (m) (f) (f)
Keysight 1250-0781 or Pomona
3285
Blocking capacitor
Keysight
Pomona 4288 +
Pomona 5088
Содержание DSOX1102A
Страница 1: ...Keysight InfiniiVision 1000 X Series Oscilloscopes Service Guide ...
Страница 7: ...1000 X Series Oscilloscopes Service Guide 7 Contents Index ...
Страница 8: ...8 1000 X Series Oscilloscopes Service Guide Contents ...
Страница 10: ...10 1000 X Series Oscilloscopes Service Guide Figures ...
Страница 12: ...12 1000 X Series Oscilloscopes Service Guide Tables ...
Страница 52: ...52 1000 X Series Oscilloscopes Service Guide 5 Replacing Assemblies ...
Страница 54: ...54 1000 X Series Oscilloscopes Service Guide 6 Replaceable Parts ...
Страница 60: ...60 1000 X Series Oscilloscopes Service Guide Index ...