344
Keysight InfiniiVision 2000 X-Series Oscilloscopes User's Guide
24
I2C/SPI Triggering and Serial Decode
3
Press the
Mode
softkey; then, select
I2C
trigger type.
4
Press the
Signals
softkey to open the I
2
C Signals Menu.
5
For both the SCL (serial clock) and SDA (serial data) signals:
a
Connect an oscilloscope channel to the signal in the device under test.
b
Press the
SCL
or
SDA
softkey; then, turn the Entry knob to select the channel
for the signal.
c
Press the corresponding
Threshold
softkey; then, turn the Entry knob to select
the signal threshold voltage level.
The threshold voltage level is used in decoding, and it will become the
trigger level when the trigger type is set to the selected serial decode slot.
Data must be stable during the whole high clock cycle or it will be
interpreted as a start or stop condition (data transitioning while the clock is
high).
The SCL and SDA labels for the source channels are automatically set.
I2C Triggering
To set up the oscilloscope to capture I2C signals, see
After the oscilloscope has been set up to capture I2C signals, you can trigger on a
stop/start condition, a restart, a missing acknowledge, an EEPROM data read, or
on a read/write frame with a specific device address and data value.
1
Press
[Trigger]
; then, select the
I2C
trigger type.
2
Press
[Trigger]
.
3
In the Trigger Menu, press the
Trigger
softkey; then, turn the Entry knob to
select the serial slot (Serial 1) on which the I
2
C signals are being decoded.
Содержание DSO-X 2014A
Страница 1: ...Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide...
Страница 8: ...8 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide...
Страница 60: ...60 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 2 Horizontal Controls...
Страница 68: ...68 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 3 Vertical Controls...
Страница 118: ...118 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 6 Digital Channels...
Страница 124: ...124 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 7 Serial Decode...
Страница 130: ...130 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 8 Display Settings...
Страница 168: ...168 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 10 Triggers...
Страница 176: ...176 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 11 Trigger Mode Coupling...
Страница 202: ...202 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 13 Cursors...
Страница 226: ...226 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 14 Measurements...
Страница 240: ...240 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 15 Mask Testing...
Страница 244: ...244 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 16 Digital Voltmeter...
Страница 256: ...256 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 17 Waveform Generator...
Страница 268: ...268 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 18 Save Recall Setups Screens Data...
Страница 274: ...274 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 19 Print Screens...
Страница 308: ...308 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 21 Web Interface...
Страница 326: ...326 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 22 Reference...
Страница 370: ...370 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 25 UART RS232 Triggering and Serial Decode...