DC RMS
DC RMS is the root-mean-square value of the waveform over one or more full
periods.
Where x
i
= value at
i
th point being measured, n = number of points in
measurement interval.
The Full Screen measurement interval variation measures the value on all
displayed data points.
The N Cycles measurement interval variation measures the value on an integral
number of periods of the displayed signal. If less than three edges are present, the
measurement shows "No edges".
The X cursors show the interval of the waveform being measured.
AC RMS
AC RMS is the root-mean-square value of the waveform, with the DC component
removed. It is useful, for example, for measuring power supply noise.
The N Cycles measurement interval measures the value on an integral number of
periods of the displayed signal. If less than three edges are present, the
measurement shows "No edges".
The X cursors show the interval of the waveform being measured.
The Full Screen (Std Deviation) measurement interval variation is an RMS
measurement across the full screen with the DC component removed. It shows the
standard deviation of the displayed voltage values.
The standard deviation of a measurement is the amount that a measurement
varies from the mean value. The Mean value of a measurement is the statistical
average of the measurement.
RMS (dc) =
∑
i
=1
n
x
i
2
n
Содержание DSO-X 2014A
Страница 1: ...Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide...
Страница 8: ...8 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide...
Страница 60: ...60 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 2 Horizontal Controls...
Страница 68: ...68 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 3 Vertical Controls...
Страница 118: ...118 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 6 Digital Channels...
Страница 124: ...124 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 7 Serial Decode...
Страница 130: ...130 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 8 Display Settings...
Страница 168: ...168 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 10 Triggers...
Страница 176: ...176 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 11 Trigger Mode Coupling...
Страница 202: ...202 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 13 Cursors...
Страница 226: ...226 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 14 Measurements...
Страница 240: ...240 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 15 Mask Testing...
Страница 244: ...244 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 16 Digital Voltmeter...
Страница 256: ...256 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 17 Waveform Generator...
Страница 268: ...268 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 18 Save Recall Setups Screens Data...
Страница 274: ...274 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 19 Print Screens...
Страница 308: ...308 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 21 Web Interface...
Страница 326: ...326 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 22 Reference...
Страница 370: ...370 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 25 UART RS232 Triggering and Serial Decode...