I2C/SPI Triggering and Serial Decode
24
Keysight InfiniiVision 2000 X-Series Oscilloscopes User's Guide
325
2
In the Search Menu, press the
Search
softkey; then, turn the Entry knob to select
the serial slot (Serial 1 or Serial 2) on which the I2C signal is being decoded.
3
Press
Search
; then, select from these options:
•
Missing Acknowledge
— finds SDA data is high during any Ack SCL clock bit.
•
Address with no Ack
— finds when the acknowledge of the selected address
field is false. The R/W bit is ignored.
•
Restart
— finds when another start condition occurs before a stop condition.
•
EEPROM Data Read
— finds EEPROM control byte value 1010xxx on the SDA
line, followed by a Read bit and an Ack bit. It then looks for the data value
and qualifier set by the Data is softkey and the Data softkeys.
•
Frame(Start:Address7:Read:Ack:Data)
— finds a read frame on the 17th clock
edge if all bits in the pattern match.
•
Frame(Start:Address7:Write:Ack:Data)
— finds a write frame on the 17th clock
edge if all bits in the pattern match.
•
Frame(Start:Address7:Read:Ack:Data:Ack:Data2)
— finds a read frame on the 26th
clock edge if all bits in the pattern match.
•
Frame(Start:Address7:Write:Ack:Data:Ack:Data2)
— finds a write frame on the
26th clock edge if all bits in the pattern match.
For more information on searching data, see
For more information on using the
[Navigate]
key and controls, see
Setup for SPI Signals
Serial Peripheral Interface (SPI) signals setup consists of connecting the
oscilloscope to a clock, MOSI data, MISO data, and framing signal, then setting
the threshold voltage level for each input channel, and finally specifying any other
signal parameters.
To set up the oscilloscope to capture SPI signals, use the
Signals
softkey which
appears in the Serial Decode Menu:
1
Press
[Label]
to turn on labels.
2
Press
[Serial]
.
3
Press the
Mode
softkey; then, select
SPI
trigger type.
Содержание DSO-X 2002A
Страница 1: ...Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide...
Страница 8: ...8 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide...
Страница 22: ...22 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide...
Страница 62: ...62 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 2 Horizontal Controls...
Страница 70: ...70 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 3 Vertical Controls...
Страница 84: ...84 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 4 Math Waveforms...
Страница 106: ...106 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 6 Digital Channels...
Страница 112: ...112 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 7 Serial Decode...
Страница 118: ...118 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 8 Display Settings...
Страница 146: ...146 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 10 Triggers...
Страница 154: ...154 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 11 Trigger Mode Coupling...
Страница 180: ...180 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 13 Cursors...
Страница 216: ...216 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 16 Digital Voltmeter...
Страница 228: ...228 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 17 Waveform Generator...
Страница 240: ...240 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 18 Save Recall Setups Screens Data...
Страница 246: ...246 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 19 Print Screens...
Страница 266: ...266 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 20 Utility Settings...
Страница 344: ...344 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 25 UART RS232 Triggering and Serial Decode...
Страница 352: ...352 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide Index...