Triggers
Pattern Trigger
The Pattern trigger identifies a trigger condition by looking for a specified pattern.
This pattern is a logical AND combination of the channels. Each channel can have
a value of 0 (low), 1 (high), or don't care (X). A rising or falling edge can be
specified for one channel included in the pattern. You can also trigger on a hex
bus value as described on
1
Press the
[Trigger]
key.
2
In the Trigger Menu, press the
Trigger
softkey; then, turn the Entry knob to
select
Pattern
.
3
For each analog or digital channel you want to include in the desired pattern,
press the
Channel
softkey to select the channel.
This is the channel source for the 0, 1, X, or edge condition. As you press the
Channel
softkey (or rotate the Entry knob), the channel you select is highlighted
in the Pattern = line directly above the softkeys and in the upper-right corner of
the display next to "Pat".
Adjust the trigger level for the selected analog channel by turning the Trigger
Level knob. Press the
[Digital]
key and select
Thresholds
to set the threshold level
for digital channels. The value of the trigger level or digital threshold is
displayed in the upper-right corner of the display.
4
For each channel you select, press the
Pattern
softkey; then, turn the Entry knob
to set the condition for that channel in the pattern.
Содержание DSO-X 2002A
Страница 1: ...Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide...
Страница 8: ...8 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide...
Страница 22: ...22 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide...
Страница 62: ...62 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 2 Horizontal Controls...
Страница 70: ...70 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 3 Vertical Controls...
Страница 84: ...84 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 4 Math Waveforms...
Страница 106: ...106 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 6 Digital Channels...
Страница 112: ...112 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 7 Serial Decode...
Страница 118: ...118 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 8 Display Settings...
Страница 146: ...146 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 10 Triggers...
Страница 154: ...154 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 11 Trigger Mode Coupling...
Страница 180: ...180 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 13 Cursors...
Страница 216: ...216 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 16 Digital Voltmeter...
Страница 228: ...228 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 17 Waveform Generator...
Страница 240: ...240 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 18 Save Recall Setups Screens Data...
Страница 246: ...246 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 19 Print Screens...
Страница 266: ...266 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 20 Utility Settings...
Страница 344: ...344 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 25 UART RS232 Triggering and Serial Decode...
Страница 352: ...352 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide Index...