Setting up and Using DP0010A/12A/13A Probes
3
DP0010A/11A/12A/13A Differential Active Probes User’s Guide
23
Configuring Probe Settings on the Oscilloscope Software GUI
When you connect your probe to the oscilloscope, the probe is automatically
detected and displayed as connected to the oscilloscope’s channel in the
Infiniium/InfiniiVision software GUI.
Figure 9
DP0011A probe auto-detected on the Infiniium GUI
Perform the following tasks on your oscilloscope’s software GUI to get accurate
measurement results.
• Configure offset behavior
• Calibrate your probe
Configuring Offset Behavior
You can configure the offset behavior to ensure that you get the maximum
performance and dynamic range from your probe. By applying an offset, most or
all of the DC component can be subtracted and the signal can be positioned to
better utilize the input’s available dynamic range.
The two offset behavior options are described and compared in the table below.