98
Keysight CX3300 User’s Guide, Edition 5
Performing User Calibration
Using the CX1104A
Figure 3-5
Connection Diagram for User Calibration (CX1104A)
After the channel passed the user calibration, the ON/OFF switch will be available.
•
Set ON to enable error correction.
Error correction is performed for the data measured by this channel from now
on.
•
Set OFF to disable error correction.
Error correction is not performed.
If the channel failed the user calibration, the ON/OFF switch is not available. Error
correction is not performed.
Once you disconnect the CX1104A from the CX3300, the user calibration data for
this channel will be cleared.
CX1104A
Extension cable
Resistive sensor head
(CX1211A
CX1212A,
CX1213A,
CX1214A
CX1215A
RU
CX1216A)
CX3300