Performing User Calibration
Using the CX1105A
Figure 3-6
Connection Diagram for User Calibration (CX1105A)
After the channel passed the user calibration, the ON/OFF switch will be available.
•
Set ON to enable error correction.
Error correction is performed for the data measured by this channel from now
on.
•
Set OFF to disable error correction.
Error correction is not performed.
If the channel failed the user calibration, the ON/OFF switch is not available. Error
correction is not performed.
Once you disconnect the CX1105A from the CX3300, the user calibration data for
this channel will be cleared.
CX1105A
Test adapter
CX3300
Slide switch can be set
on either side.
Leave the BNC
connector open.
KEYSIGHT
SIGNAL TO
POS AND NEG
CX1105A
Test adapter
Insert