Keysight B2900 User’s Guide, Edition 5
4-39
Front Panel Reference
Trigger key group
Trigger Output
Trigger output, ON or OFF
Trigger output terminal, an internal bus (INT1, INT2), LAN
port (LAN), or a digital I/O pin EXT
n
(
n
=1 to 14)
If this parameter is ON, B2900 sends an output trigger when it
changes the trigger state for the action specified by the Ch,
Layer, and Action parameters. See
For using a digital I/O pin, it is necessary to set function of the
pin, output trigger polarity, type, pulse width, and so on. See
“DIO Configuration dialog box” on page 4-48
Table 4-8
Trigger Source
Trigger Source
Description
AUTO
Trigger source best suited for the present operating mode is
automatically selected by the internal algorithms.
BUS
Remote interface trigger command such as GET, TRIGger,
and *TRG
TIMER
Signal internally generated every interval set by the Period
field
INT1 or INT2
Signal from the internal bus 1 or 2
LAN
LXI trigger specified by the
:ARM[:ACQ|:TRAN]:SOUR:LAN and
:TRIG[:ACQ|:TRAN]:SOUR:LAN commands
EXT
n
Signal from the DIO pin
n
, which is an I/O port of the Digital
I/O D-sub connector on the rear panel.
n
=1 to 14
Содержание B2900 Series
Страница 1: ...Keysight Technologies B2900 Series Precision Source Measure Unit User s Guide...
Страница 3: ......
Страница 25: ...1 Getting Started...
Страница 41: ...2 Introduction...
Страница 64: ...2 24 Keysight B2900 User s Guide Edition 5 Introduction Options...
Страница 65: ...3 Installation...
Страница 101: ...4 Front Panel Reference...
Страница 155: ...5 Front Panel Operations...
Страница 189: ...6 Function Details...
Страница 233: ......