3-3
Performing System Setup and Compensation
Pulsed IV System Setup
Pulsed IV System Setup
The pulsed IV system setup items are listed below. The system setup must be
performed and successfully completed before using the pulsed IV test system.
Accessories needed to perform the system setup are listed in
. Prepare the
accessories additionally to the system accessories and the system components.
•
System Configuration
This action is necessary to specify the interface address and the source or
measurement channels of PGU (pulse generator), DSO (digital sampling
oscilloscope), and SMU (source monitor unit) used for the pulsed IV
measurement. See
“System Configuration” on page 3-10
. For the setup
parameters, see
.
•
Compensation
This action is necessary to obtain and update the compensation factors of the
pulsed IV test system.
•
Skew Measurement
This action performs the skew measurement which is one of the measurements
performed in the Compensation action. If the skew measurement error occurs
during the Compensation action is executed, you can perform only the skew
measurement again by this action.
•
Drain Cable Replacement
This action is necessary to obtain and update the compensation factors of the
bias-T and the cable of the drain measurement path.
•
Pgu Compensation
This action obtains the compensation factors of the pulse generator. This action
should be executed before you perform the pulsed IV measurements which set
the gate voltage automatic adjustment (
VgAdjust
) to off (
Disable
).
When you must perform the system setup is shown in
.
To execute an action, see
“Executing Action” on page 3-12
.
Содержание B1542A Pulsed IV Package
Страница 1: ...Keysight Technologies B1542A Pulsed IV Package for B1500A EasyEXPERT User s Guide ...
Страница 9: ...1 Introduction ...
Страница 18: ...1 10 Keysight B1542A User s Guide Edition 10 Introduction Accessories and Options ...
Страница 19: ...2 Installation ...
Страница 41: ...3 Performing System Setup and Compensation ...
Страница 54: ...3 14 Keysight B1542A User s Guide Edition 10 Performing System Setup and Compensation Executing Action ...
Страница 55: ...4 Performing Measurement ...
Страница 68: ...4 14 Keysight B1542A User s Guide Edition 10 Performing Measurement DC I V Measurements ...
Страница 69: ...5 PLSDIV Test Definitions ...
Страница 108: ...5 40 Keysight B1542A User s Guide Edition 10 PLSDIV Test Definitions Parameters ...
Страница 109: ...6 PLSDIV TIS Commands ...
Страница 133: ...7 Status Code and Error Messages ...
Страница 159: ......