Keysight B1507A User’s Guide, Edition 1
2- 13
Installation
Inspection and Installation
To Check Operation of Selector
1. Complete the cable connections described in “To Connect Selector and Test
2. Connect the power cable from the selector to an AC power outlet.
3. Connect the power cable from the mainframe to an AC power outlet.
4. Open the measurement terminals of the test fixture and close the fixture cover.
5. Set the selector power switch to the ON position to turn on the selector.
6. Press the mainframe Standby switch to turn on the mainframe.
7. After boot up, launch the Capacitance Measurement software.
8. Click the Configuration > Module menu to display the Module Configuration
dialog box.
9. Perform diagnosis of the selector by using the Module Configuration dialog box.
If problems arise, contact Keysight Technologies.
To Change Windows Logon Setting
Windows automatic logon has been enabled initially. If you want to change this
setting, select
All Programs
>
Control Auto Logon
from the Start menu. Control
Auto Logon is displayed on the browser. Then click one of the following buttons on
the browser and follow the setup script.
•
Enable Auto Logon
•
Disable Auto Logon
NOTE
If you set a password for the user account
Keysight B1500 User
, disable the
automatic logon.