CALibrate Subsystem
This subsystem calibrates the instrument.
Follow the procedure in the calibration section before calibrating. Improper
calibration can reduce accuracy and reliability.
CALibrate:COUNt?
Returns the calibration count.
Parameter
Typical Return
(none)
+14
Return the calibration count: CAL:COUN?
The calibration count is incremented whenever any of the following commands is executed:
– CALibrate:SAVE ([Menu] > System > Admin > Cal > Save)
– CALibrate:DATE ([Menu] > System > Admin > Cal > Date)
– CALibrate:PASSword ([Menu] > System > Admin > Password)
– CALibrate:PASSword:RESet
You may execute this query regardless of the CALibrate:STATe value.
CALibrate:CURRent:OFFSet <
value
>
Starts the DC current limit calibration for the specified voltage range.
Parameter
Typical Return
The maximum current of the output range being calibrated, as shown in the table below. (none)
Calibrate the 10-A current range: CAL:CURR:OFFS 10
The
must be ON to execute this command.
The <
value
> that you specify will select the range to be calibrated, as shown in the table below. For
example, a <
value
> of 8 will select the 310 V range on the AC6803B, but the 155 V range on the
AC6802B.
AC6801B AC6802B AC6803B AC6804B
310 V Range
2.0
4.0
8.0
16.0
155 V Range
4.0
8.0
16.0
32.0
Do not use any unit suffix (such as A) in this command.
3 SCPI Programming Reference
116
Keysight AC6800B Series Operating and Service Guide
Содержание AC6801B
Страница 1: ...Operating and Service Guide Keysight AC6800B Series Basic AC Sources...
Страница 16: ...AC6802B 1 Getting Started 16 Keysight AC6800B Series Operating and Service Guide...
Страница 17: ...AC6803B Keysight AC6800B Series Operating and Service Guide 17 1 Getting Started...
Страница 18: ...AC6804B 1 Getting Started 18 Keysight AC6800B Series Operating and Service Guide...
Страница 62: ......
Страница 196: ...4 Calibration Verification and Service 196 Keysight AC6800B Series Operating and Service Guide...