![Keysight Technologies 85091D Скачать руководство пользователя страница 115](http://html.mh-extra.com/html/keysight-technologies/85091d/85091d_reference-manual_1961832115.webp)
Electronic Calibration Modules Reference Guide 5-9
Specifications and Characteristics
Electrical Specifications
N4431D (7-16)
The electrical performance in the following table applies to N4431D (if any of
the ports consist of 7-16 connector).
N4431D (4.3-10)
The electrical performance in the following table applies to N4431D (if any of
the ports consist of 4.3-10 connector).
Table 5-14 Electrical Performance for N4431D (m/f 7–16 connectors)
a
Parameter
Frequency Range
DC to 10 MHz
10 MHz to 5 GHz 5 GHz to 7.5 GHz
Directivity (dB)
45
50
47
Source Match (dB) 36
42
39
Reflection Tracking
(±dB)
0.10
0.06
0.09
a. When applied power exceeds –7 dBm, calibration results will be degraded from the performance
indicated in this table.
Table 5-15 Electrical Performance for N4431D (m/f 4.3–10 connectors)
a
Parameter
Frequency Range
DC to 10 MHz
10 MHz to 5 GHz 3 GHz to 9 GHz
9 GHz to 12 GHz
Directivity (dB)
45
50
47
35
Source Match (dB) 36
42
39
37
Reflection Tracking
(±dB)
0.10
0.06
0.09
0.10
a. When applied power exceeds –7 dBm, calibration results will be degraded from the performance
indicated in this table.
Содержание 85091D
Страница 4: ...4 ...
Страница 44: ...1 36 Electronic Calibration Modules Reference Guide General Information Safety and Regulatory Information 1 ...
Страница 66: ...2 22 Electronic Calibration Modules Reference Guide Preparing ECal for Use Operating and Safety Precautions 2 ...
Страница 82: ...3 16 Electronic Calibration Modules Reference Guide Operating ECal Module ECal Troubleshooting 3 ...
Страница 170: ...6 14 Electronic Calibration Modules Reference Guide Replaceable Parts Other ECal Accessories 6 ...
Страница 171: ......