5-4 Electronic Calibration Modules Reference Guide
Specifications and Characteristics
Electrical Specifications
5-
Electrical Specifications
Electrical performance for RF and microwave ECal modules are provided in the
following tables, which describe warranted performance that most units
exhibit.
8509xD Series
ECal modules reflection parameter performance applies to each of the
ports.
Table 5-3 Electrical Performance for 85091D (7 mm)
Parameter
Frequency Range
DC to
10 MHz
a
a. For option 0DC.
300 kHz
to 10 MHz
b
b. For option 003.
10 MHz to
1 GHz
1 to 3 GHz
3 to 6 GHz
6 to 9 GHz
Directivity (dB)
52
45
52
52
50
45
Source Match (dB)
45
36
45
44
41
34
Reflection Tracking (±dB)
0.04
0.10
0.04
0.04
0.07
0.10
Table 5-4 Electrical Performance for 85092D (Type-N 50
Ω
)
Parameter
Frequency Range
DC to
10 MHz
a
a. For option 0DC.
300 kHz
to 10 MHz
b
b. For option 003.
10 MHz to
1 GHz
1 to 3 GHz
3 to 6 GHz
6 to 9 GHz
Directivity (dB)
52
45
52
52
49
45
Source Match (dB)
45
36
45
44
41
36
Reflection Tracking (±dB)
0.04
0.10
0.04
0.04
0.06
0.07
Содержание 85091D
Страница 4: ...4 ...
Страница 44: ...1 36 Electronic Calibration Modules Reference Guide General Information Safety and Regulatory Information 1 ...
Страница 66: ...2 22 Electronic Calibration Modules Reference Guide Preparing ECal for Use Operating and Safety Precautions 2 ...
Страница 82: ...3 16 Electronic Calibration Modules Reference Guide Operating ECal Module ECal Troubleshooting 3 ...
Страница 170: ...6 14 Electronic Calibration Modules Reference Guide Replaceable Parts Other ECal Accessories 6 ...
Страница 171: ......