Keysight 85059V Operation and Service Guide 3-13
Use, Maintenance, and Care of the Devices
How to Make a Connection
Figure 3-6
Alignment
Connector Misalignment
Forced
misalignment could damage the female center conductor.
Figure 3-7
Misalignment
How to Separate a Connection
To avoid lateral (bending) force on the connector mating plane surfaces,
always support the devices and connections.
1.
Use an open-end wrench to prevent the device body from turning.
Turn the connector nut,
not
the device body. Major damage to the center
conductor can occur if the device body is twisted.