1-8 Keysight 85059B Operation and Service Guide
General Information
Calibration Residual Error Specifications
1-
Calibration Residual Error Specifications
Please download our free Vector Network Analyzer Uncertainty Calculator from
http://www.keysight.com/find/na_calculator
.
Table 1-1
85059B (Precision) Residual System Calibration Error
Specifications
Frequency
Range (GHz)
Data-Based Model (Residuals)
1
“Data-Based Model for Defining the
Calibration Standards” on page 1-3
.
Directivity
Source
Match
Reflection
Tracking
DC to 10
–29
–28
0.02
10 to 15
–31
–31
0.09
15 to 20
–35
–35
0.09
20 to 30
–37
–36
0.05
30 to 40
–37
–35
0.05
40 to 50
–33
–32
0.07
50 to 60
–31
–30
0.08
60 to 70
–31
–30
0.15
70 to 80
–28
–27
0.15
80 to 100
–28
–27
0.11
100 to 115
–27
–27
0.13
115 to 120
–27
–27
0.13
Содержание 85059B
Страница 1: ...Keysight Technologies 85059B 1 0 mm Precision Calibration Kit Operating and Service Manual...
Страница 6: ...4 Keysight 85059V Operation and Service Guide Contents...
Страница 22: ...1 16 Keysight 85059B Operation and Service Guide General Information Regulatory and Environmental Information 1...
Страница 48: ...3 20 Keysight 85059B Operation and Service Guide Use Maintenance and Care of the Devices Handling and Storage 3...
Страница 66: ...85059V Index Index 4 part numbers 6 2 6 4...
Страница 67: ...Installation Note Xxxxx xxxxx 3...