85058B/E
3-7
Use, Maintenance, and Care of the Devices
Cleaning Connectors
. When cleaning a female connector, avoid snagging the swab on the center
conductor contact fingers by using short strokes.
c. Let the alcohol evaporate, then blow the connector dry with a gentle stream of clean,
low-pressure compressed air or nitrogen. Always completely dry a connector before you
reassemble or use it.
4.
Reinspect
Inspect the connector again to make sure that no particles or residue are present.