Performance Tests
1
Keysight 53210A/53220A/53230A Assembly Level Service Guide
37
Keysight 53210A/53220A/53230A Complete Performance Tests
The following seven performance tests, when successful, verify that the counter
meets the instrument specifications as provided on the CD ROM that came with
your Keysight 53210A/53220A/53230A Counter. See also the information on
running a ‘QUICK’ performance test on
. Before running these tests,
review the
Record the results of the performance tests in the appropriate place on the
“Keysight 53210A/53220A/53230A Performance Test Record (Tests 1 to 7)”
page 90.
To verify the total number of accuracy places for frequency measurements, it may
be necessary to turn off Auto-Digit control. Use this set of key depressions to turn
AutoDigits OFF:
[Digits] / (AutoDigits) / <Off>
NOTE
If the instrument is currently located in the calibration environment with the
power removed, turn on the instrument and allow a 45 minute warm-up period
before starting the Performance Test procedures. Upon receipt and delivery to
the calibration environment, turn on the instrument and allow a 2½ hour
warm-up/stabilization period before starting the procedures.
If it is known that the instrument has been in storage or decommissioned
(unused) for several months, turn on the instrument and allow it to stabilize for
72 hours. Periodically monitor the aging (drift) rate relative to the counter’s
24-hour specification. If the aging rate exceeds the 24-hour specification, allow
an additional stabilization period and continue to monitor the rate relative to the
specification. The counter has stabilized when aging remains within the 24-hour
rate. Otherwise, the 45 minute or 2½ warm-up period applies.
An instrument that has not or will not be used for several months should be
stored in the same ambient environment and orientation in which it will
eventually be used.
Содержание 53210A
Страница 1: ...Keysight 53210A 53220A 53230A 350 MHz Universal Frequency Counter Timer Assembly Level Service Guide ...
Страница 18: ...18 Keysight 53210A 53220A 53230A Assembly Level Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 22: ...22 Keysight 53210A 53220A 53230A Assembly Level Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 180: ...2 Service 180 Keysight 53210A 53220A 53230A Assembly Level Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...