background image

Содержание Riceter fg Series

Страница 1: ...Grain Rice Moisture Tester Riceter fg series Operating Manual...

Страница 2: ...s This tester can measure the moisture of grain by simple operation but proper operation is necessary for accurate moisture measurement Please read this operating manual carefully before use Note The...

Страница 3: ...2 Part Names 3 Description of Display 4 How to Place Batteries 6 Before Measurement 7 Measurement 8 Average Value 15 Storage 16 Rice Husker Instructions 17 Roller Assembly Replacement 21 Specification...

Страница 4: ...2 Contents of Package Operating manual Rice husker TR 130 Carrying Case Sample tray 2 pcs Spoon with tweezers Cleaning brush Battery AA sizex4 pcs Inspection Certificate...

Страница 5: ...utton SELECT button Display Application table Handle stop red mark Model No MEA button Measure AVE button Average Testing Chamber Handle Note Application table differ depending on the model number Abo...

Страница 6: ...ly turned off in approx 5 minutes after turning on the power if no operation is performed Use of backlight in the display section allows users to see the display clearly even in the dark Battery Indic...

Страница 7: ...ell below the lower limit of the measurement range or the contact between the test sample tray and the measuring unit of the main body was poor Error mark Main unit temperature is too low 5 C or lower...

Страница 8: ...ke out the main unit from the carrying case Take out the batteries from under the battery holder Open the battery box on the back side of the main unit and place 4 batteries appropriately How to Place...

Страница 9: ...re gradually discharging even when the tester is not used It is recommended to always use new batteries Place batteries in the correct polarity Riceter fg500 series is equipped with a temperature comp...

Страница 10: ...er Press the POWER button All of the backlight all characters on the display section mark light up for 2 seconds and then model number and software version number is displayed automatically The backli...

Страница 11: ...cation Times are displayed It is ready for measurement Measurement Press the SELECT button and repeat until the Selection Indicator and the abbreviation for name of application you wish to test appear...

Страница 12: ...will be one layer Note Both the amount of sample is too much or insufficient may cause a measurement error Over more too much sample may cause wear on the screw and damage the instrument Note Sort ou...

Страница 13: ...in the sample tray and insert it into the testing chamber until the guide line is hid Note Failure to observe this may cause the tip of the handle to damage the plastic portion of the sample tray Be s...

Страница 14: ...e MEA button the decimal point blinks the backlight lights up the moisture content and measuring number are displayed thereafter The backlight goes out after 4 seconds but the moisture content remains...

Страница 15: ...e is displayed The measuring count alsochangesatthesametime 7 Measurement After a lapse of 5 minutes the power is automatically turns off and all the displayed items are cleared Pressing the POWER but...

Страница 16: ...r every measurement If the last test sample remains proper measurement cannot be performed For continuous measurement disconnect the handle and remove the attached sample and contamination from the ti...

Страница 17: ...btain the average value of m o i s tu r e co nte nt s p r e ss th e AVE button within 5 minutes after measurement After the last measured value disappears the average calculation function does not wor...

Страница 18: ...lean every portion of the main body in a careful manner Especially for the measuring unit disconnect the handle and clean the inner contact section sufficiently Be sure to put the tester together with...

Страница 19: ...17 Rice Husker Instructions Rice Husker TR 130 Parts Paddy hopper Brown rice outlet Handle Husk box Attach the handle to the body 1...

Страница 20: ...18 Rice Husker Instructions Pour the paddy into the hopper and close the cover While holding the TR 130 at the edge of a desk or table turn the handle in the clockwise direction to husk the sample 2 3...

Страница 21: ...rice kernels may be broken by husking Recommended Method There are two methods to remove the brown rice 1 Open the top of the paddy hopper Turn the instrument over and the brown rice will be removed...

Страница 22: ...20 Rice Husker Instructions Discard the husk before the husk box is full...

Страница 23: ...n the rollers are worn to maintain husking efficiency Also replace the assembly when foreign objects have been caught in the rollers Turn the main unit upside down Remove the lower case by removing th...

Страница 24: ...22 Roller Assembly Replacement Mount the new roller assembly and confirm the gears mesh Assemble in the reverse order of disassembly If the handle does not turn well please try again 3 4...

Страница 25: ...ction is applicable less than 20 moisture range of sample Power source 1 5V AA size battery x 4 Auto power off 5 min after nothing of operation Power consumption Max 0 3W Dimensions and weight 164 W x...

Страница 26: ......

Страница 27: ......

Страница 28: ...ted in this user manual may not ap pear exactly as their actual counterparts but this does not affect their operation or functionality This user manual was intended to be written as clearly and accura...

Отзывы: