PNJ_PNS-BA-e-1612
31
7.3 Zeroing
In order to obtain optimal weighing results, reset to zero the balance before weighing.
Unload the balance.
Press the
TARE
key.
Wait until the zero display and the indicator a will
appear.
During the zeroing procedure a flashing „M“ will be displayed.
7.4 Simple weighing
Place goods to be weighed on balance.
Wait for stability display
.
Read weighing result.
Capacity display
[
]
At active capacity display (see chap. 8.1.2 „1.b.G.1“) the bar graph moves
from the left to the right and proceeds equally to the weight loaded onto the
weighing balance. Its full width is reached at maximum load. This is an
analogue display of the current allocation of the weighing area.
Overload warning
Overloading exceeding the stated maximum load (max) of the device,
minus a possibly existing tare load, must be strictly avoided. This could
damage the instrument.
Exceeding maximum loads is indicated by the display of “o-Err“, and an
audio sound. Unload weighing system or reduce preload.