AEJ-C/AES-C/PLJ-C-BA-e-1621
59
13.1 Automatic adjustment using internal weight (models AEJ, PLJ)
With the internal adjustment weight, the weighing accuracy can be checked and re-
adjusted at any time.
The automatic adjustment function is always enabled. You can start adjustment at
any time by pressing the
CAL
-key manually.
Temperature / time controlled adjustment is taking place:
Two minutes prior to automatic adjustment „CAL“ in combination with a clock and
thermometer icon will be shown at the upper edge of the display.
or
The user must complete his/her weighing process within this time. After two minutes
[
CAL 30 ]
appears on the display.
This starts a “count down” of 30 seconds
[ CAL 30 ]
→
[ CAL 0 ].
You may cancel
adjustment during these 30 seconds by pressing the
TARE
-key. This makes the
balance return to weighing mode in order to e. g. complete an unfinished
measurement.
After approx. further 5 minutes the automatic adjustment will restart and
[ CAL 30 ]
appears.
You may interrupt the adjustment process several times. However, keep in mind that
an interruption of the adjustment may lead to incorrect weighing results.
The adjustment log will be printed automatically if an optional printer is connected.
For defining an adjustment log see chap. 14.1 / <P5.1 Adjustment log>.
Printout example KERN YKB-01N:
Содержание PLJ 300-3CM
Страница 11: ...AEJ C AES C PLJ C BA d 1621 10 Abmessungen AEJ C AES C PLJ C...
Страница 21: ...AEJ C AES C PLJ C BA d 1621 20...
Страница 49: ...AEJ C AES C PLJ C BA d 1621 48...
Страница 81: ...AEJ C AES C PLJ C BA d 1621 80 15 1 2 Benutzer l schen...
Страница 83: ...AEJ C AES C PLJ C BA d 1621 82 15 2 1 Neues Produkt anlegen Produktbezeichnung festlegen...
Страница 84: ...83 AEJ C AES C PLJ C BA d 1621 Nach Eingabe der Produktbezeichnung nacheinander alle Parameter definieren...
Страница 113: ...AEJ C AES C PLJ C BA d 1621 112...
Страница 115: ...AEJ C AES C PLJ C BA d 1621 114...
Страница 130: ...129 AEJ C AES C PLJ C BA d 1621 26 2 Durchf hrung Applikation anw hlen...
Страница 138: ...137 AEJ C AES C PLJ C BA d 1621 Nach Auswahl der Baudrate n chsten Kommunikationsparameter Parit t w hlen...
Страница 139: ...AEJ C AES C PLJ C BA d 1621 138 27 2 P3 3 WLAN Schnittstelle Factory option KERN PLJ A06 Men aufrufen...
Страница 173: ......
Страница 182: ...10 AEJ C AES C PLJ C BA e 1621 Dimensions AEJ C AES C PLJ C...
Страница 192: ...20 AEJ C AES C PLJ C BA e 1621...
Страница 220: ...48 AEJ C AES C PLJ C BA e 1621...
Страница 252: ...80 AEJ C AES C PLJ C BA e 1621 15 1 2 Delete user...
Страница 254: ...82 AEJ C AES C PLJ C BA e 1621 15 2 1 Create new product Define product name...
Страница 255: ...AEJ C AES C PLJ C BA e 1621 83 After entering a product name define all parameters one by one...
Страница 284: ...112 AEJ C AES C PLJ C BA e 1621...
Страница 286: ...114 AEJ C AES C PLJ C BA e 1621...
Страница 301: ...AEJ C AES C PLJ C BA e 1621 129 Select application...
Страница 309: ...AEJ C AES C PLJ C BA e 1621 137 After selecting baud rate select the next communication parameter Parity...
Страница 310: ...138 AEJ C AES C PLJ C BA e 1621 27 2 P3 3 WLAN interface factory option KERN PLJ A06 Call up menu...
Страница 352: ...10 AEJ C AES C PLJ C BA f 1621 Dimensions AEJ C AES C PLJ C...
Страница 362: ...20 AEJ C AES C PLJ C BA f 1621...
Страница 390: ...48 AEJ C AES C PLJ C BA f 1621...
Страница 422: ...80 AEJ C AES C PLJ C BA f 1621 15 1 2 Effacement de l usager...
Страница 424: ...82 AEJ C AES C PLJ C BA f 1621 15 2 1 Constituer un nouveau produit D finir une d signation de produit...
Страница 425: ...AEJ C AES C PLJ C BA f 1621 83 Apr s saisie de la d signation du produit d finir successivement tous les param tres...
Страница 454: ...112 AEJ C AES C PLJ C BA f 1621...
Страница 456: ...114 AEJ C AES C PLJ C BA f 1621...
Страница 471: ...AEJ C AES C PLJ C BA f 1621 129 26 2 R alisation S lectionner l application...
Страница 480: ...138 AEJ C AES C PLJ C BA f 1621 27 2 P3 3 interface WLAN Factory option KERN PLJ A06 Appel du menu...