The display for weight determination of the "glass sinker in air“ appears.
Confirm by pressing the
button.
Tare if necessary and hook on the glass sinker centrically.
Wait until the weight display of the weighing balance has become stable and
then press the
button. „WEI AIR“ appears flashing.
The display for weight determination of the "glass sinker in sample liquid“
appears.
Acknowledge using
button
Tare if necessary and immerse the sinker into the measuring fluid
Wait until the weight display of the weighing balance has become stable and
then press the
button. „WEI LIQ“ appears flashing.
The density of the sample fluid is shown
Return to weighing mode
Press the
ON/OFF
key
or use the
MENU
button to start a new measuring cycle.
If at the density determination errors have appeared, „d-----„ is displayed.